期刊文献+

基于ARM的人体表面温度测控系统设计

The Measurement and Control System of Shell Temperature of Human Body Based on ARM
下载PDF
导出
摘要 针对手术过程中人体的低温现象,开发一种基于ARM9 2440嵌入式处理器的、以人体表面温度为控制对象的表面温度控制系统。系统由多路人体表面温度检测模块、触摸式显示控制模块、气体温度检测及加热模块、气体流量检测及输送控制模块等构成。对系统核心模块的功能和效果进行试验,结果表明,在手术中可以根据人体表面温度的实时变化情况,控制吹拂在人体表面的气流温度和流量,达到使人体核心温度稳定的效果。 Aimed at the hypothermia during the medical operation a control system of shell temperature of human body was developed based on ARM9 2440 embedded processor, which consisted of several intelligence control modules such as a muhi-chanel temperature detection module, a touch sensitive display and control module, an air temperature detection and warming module, a air flux detection and control module, etc. Experiments were carried out to validate the core modules. The results show that the core temperature of human body can be maintained by controlling the temperature and air flux onto the human skin according to the instant shell temperature.
出处 《机床与液压》 北大核心 2010年第14期52-55,共4页 Machine Tool & Hydraulics
基金 2008年广西教育厅科研项目(200808LX136) 2009年广西研究生科研创新项目(2009105950802M03)
关键词 控制系统 ARM嵌入式系统 体表温度 Control system ARM embedded processor Shell temperature
  • 相关文献

参考文献4

二级参考文献17

  • 1杨帆,赵世平,黎会鹏.基于单片机和虚拟仪器的温度测控系统设计[J].武汉化工学院学报,2005,27(5):65-67. 被引量:3
  • 2华惠娟.麻醉期间体温的监测[J].国外医学:麻醉学与复苏分册,1983,3:182-185.
  • 3Fan yang, Guoping Li, Huipeng Li . Design & Development Of A Remote Temperature Monitor System Of Web Using Virtual Instruments. [J]. advances in computer, information,and systems sciences and engineering. 2006:449 - 452
  • 4Dezra Hinkson, Claude Marshall, Sham Robinson, "Design & development Of a user interface to remotely control a radio telescope using virtual instruments, " in Proc. IEEE Southeastcon 2002, pp. 279 - 282
  • 5T Sundic, S Marco, J Samitier et al. Electronic Tongue and Electronic Nose Data Fusion in Classification with Neural Networks and Fuzzy Logic Based Models[J] . IEEE Transactions on Systems Man and Cyberntetics, 1999, 23(3): 1474 - 1479
  • 6Ikeda T, Sessler D,Marder D, et al. Influence of thermoregulatory vasomotion and ambient temperature variation on the accuracy of core-temperature estimates by cutaneous liquidcrystal thermometers [ J ] . Anesthesiology, 1997,86 ( 3 ): 603~612.
  • 7Lloyd EL. Accidental hypothermia[ J]. Resuscitation, 1996,32(2) :111 - 1124.
  • 8Frank SM, Fleisher LA, Breslow MJ, et al. Perioperative maintenance of normothermia reduces the incidence of morbid cardiac events. Arandomized clinical trial [ J ]. JAMA,1997,277(14) :1127 - 1134.
  • 9Beilin B, Shavit Y, Razumovsky J, et al. Effects of mild perioperative hypothermia on cellular immune responses [ J ].Anesthesiology, 1998,89 ( 5 ): 1133 - 1140.
  • 10Schmied H,Kurz A,Sessler DI,et al.Mild hypothermia increases blood loss and transfusion requirements during total hip arthroplasty [ J ]. Lancet, 1996,3: 347 ( 8997 ): 289 -292.