3种测量石蜡比定压热容实验方法的比较

Comparison of Three Methods Measuring Heat Capacity of Paraffin
下载PDF
导出
摘要 用差示扫描法、气相色谱法和水浴法对石蜡的比定压热容进行了测量。差示扫描法可以给出石蜡比定压热容的精确值。气相色谱法通过测量各烷烃的质量分数,可较准确地对液态石蜡的比热容进行测量。相对于差示扫描实验的结果,误差小于3.9%,用这种方法测量固态石蜡的比热容存在较大误差。水浴法可对固态石蜡的比热容进行较准确的测量,误差小于3.5%。用该法测量液态石蜡的比热容时,必须对水浴上方存在液态石蜡时的散热进行准确测量。 Differential scanning calorimeter( DSC ), gas chromatography (GC) , and adiabatic bath are employed to measure heat capacity of paraffin respectively. DSC provides precise values of heat capacity for both solid and liquid paraffin. GC gives mass fractions of alkanes in paraffin, which can generate heat capacity combined with a proper mix nile. Compared with the results of liquid paraffin given by DSC, relative errors of results given by GC are less than 3.9%. Melting of iso-alkanes occur during heating of solid paraffin, that causes problems in measuring heat capacity of solid paraffin by this method. Relative errors of heat capacity of solid paraffin measured by adiabatic bath are less than 3.5%. When paraftin is at liquid phase, heat release of the bath to surroundings must be carefully calibrated.
出处 《计量学报》 CSCD 北大核心 2010年第4期308-311,共4页 Acta Metrologica Sinica
基金 基金项目:江苏省高校自然科学研究计划(06KJD480097)
关键词 计量学 比定压热容 差示扫描量热 气相色谱 绝热式水浴量热 Metrology Heat capacity Differential scanning calorimeter Gas chromatography Adiabatic bath
  • 相关文献

参考文献4

  • 1Demirel Y, Paksoy H O. Thermal Analysis of Heat Storage Materials[J]. Thermochimica Acta, 1993, 213 (14): 211 - 221.
  • 2孙建平,张金涛,邱萍,段宇宁.高精密自动绝热量热计的实验研究与分析[J].计量学报,2006,27(4):331-334. 被引量:5
  • 3Busserolles K B, et al. Calibration of a fixed-cell temperature-scanning calorimeter to measure precise solution heat capacities from 275 to 398 K at 0. 35 MPa [ J ]. Thermochimica Acta, 2000, 347 ( 1,2 ) : 3 - 7.
  • 4Ji Hong Yan, Bahman Tohidi, Ali Danesh, et al. Wax phase equilibria: developing a thermodynamic model using a systematic approach [ J ]. Fluid Phase Equilibria, 2004, 216(2) : 201 -217.

二级参考文献5

  • 1张金涛,原遵东,邱萍,段宇宁,孙建平.高精密自动绝热量热计[J].计量学报,2005,26(4):320-325. 被引量:6
  • 2Sorai M,Kaji K,Kaneko Y.An automated adiabatic calorimeter for the temperature range 13 K to 530 K-the heat capacities of benzoic acid from 15 K to 305 K and of synthetic sapphire from 60 K to 505 K[J].J Chem Thermodynamics,1992,24:167~180.
  • 3Inaba A.An adiabatic calorimeter for use at intermediate and higher temperatures-the heat capacity of synthetic sapphire (α-Al2O3) from 70 to 700 K[J].J Chem Thermodynamics,1983,15:1137~1143.
  • 4Ditmars D A,Douglas T B.Measurement of the relative enthalpy of pure α-Al2O3 (NBS heat capacity and enthalpy standard reference material No.720) from 273 to 1 173 K[J].J Res NBS,Physics and Chemistry,1971,75 A:401~420.
  • 5Archer D G.Thermodynamic properties of synthetic sapphire (α-Al2O3),standard reference material 720 and the effect of temperature-scale differences on thermodynamic properties[J].J Phys Chem RefData,1993,22:1441~1451.

共引文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部