标本放置时间对MICRO TEST1血沉仪测定结果的影响 被引量:6

Impact of erythrocyte sedimentation result detected by MICRO TEST 1 when standing time of specimen in vitro is different
下载PDF
导出
摘要 目的探讨标本放置时间对MICRO TEST1血沉仪测定结果的影响。方法抽取正常和异常组各30例血标本,放置1 h、2 h、4 h、6 h、12 h,分别用MICROTEST1血沉仪进行测定,并挑选6份标本,每份标本重复检测10次来评价血沉仪的精密度。结果 1 h与2 h测定差异无统计学意义(P1>0.05),1 h与4 h和6 h测定差异有统计学意义(P2<0.05,P3<0.05),而12 h测定结果明显低于4 h的测定值。结论 TEST1血沉仪测定血沉应在2 h内检测完毕。TEST1血沉仪的精密度比较满意,可以满足临床的检测需要。 Objective To investigate impact of erythrocyte sedimentation rate t detected by MICRO TEST 1 in different times.Methods To draw 60 blood specimens anticoagulated by EDTA-K2.There were 30 normal and 30 abnormal specimens respectively.Erythrocyte sedimentation was detected by MICROTEST1 when standing time was 1 h,2 h,4 h,6 h and 12 h.Precision was evaluated by 6 specimens detected 10 times per specimen repeatedly.Results There were no difference significantly between 1 h and 2 h but difference significantly between 1 h and 4 h,6h.Results detected at 12 h were lower significantly than that at 4 h.Conclusion Specimens for detecting ESR should be detected in 2h by MICRO TEST 1.Precision of MICRO TEST 1 is satisfied relatively.
出处 《安徽医学》 2010年第7期823-824,共2页 Anhui Medical Journal
关键词 红细胞沉降率 放置时间 血沉仪 Erythrocyte sedimentation rate Standing time Instrument for detecting erythrocyte sedimentation
  • 相关文献

参考文献2

二级参考文献10

  • 1Plebani M. Eryterocyte sedimentation rate: innovative techniques for an obsolete test? Clin Chem Lab Med, 2003, 41:115-116.
  • 2Beinhart WH. Erythrocyte sedimentation rate-more than an old fashion? Ther Umsch, 2006, 63 : 108-112.
  • 3International Committee for Standardization in Hematology.Reference method for the erythrocyte sedimentation rate ( ESR )test on human blood. Br J Haematol, 1973,24:671-673.
  • 4Fabry TL. Mechanism of Erythrocyte Aggregation and Sedimentation. Blood, 1987,70 : 1572-1576.
  • 5Plebani M, De Toni S, Sanzari MC, et al. The TEST 1 Automated System: a new method for measuring the erythrocyte dedimentation rate. Am J Clin Pathol, 1998, 110:334-340.
  • 6Piva E, Sanzari MC, Servidio G, et al. Length of sedimentation reaction in undiluted blood (erythrocyte sedimentation rate ):variations with sex and age and reference limits. Clin Chem Lab bged, 2001,39:451-454.
  • 7Piva E, Fassina P, Plebani M. Determination of the length of sedimentation reaction (erythrocyte sedimentation rate ) in nonanticoagulated blood with the Microtest 1. Clin Chem Lab Med,2002, 40:713-717.
  • 8Romero A, Munoz M, Ramirez G. Length of sedimentation reaction in blood : a comparison of the test 1 ESR system with the ICSH reference method and the sedisystem 15. Clin Chem Lab Med, 2003, 41: 232-237.
  • 9Giavarina D, Capuzzo S, Cauduro F, et al. Internal quality control for erythrocyte sedimentation rate measured by TEST-1 analyzer. Clin Lab, 2002, 48:459-462.
  • 10李晖.切忌用千篇一律的模式估计正常值范围[J].中华内科杂志,1994,33:148-149.

共引文献6

同被引文献24

引证文献6

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部