期刊文献+

XE-2100全自动血细胞分析仪计数有核红细胞的准确性评价 被引量:10

The Accuracy Evaluation of Nucleated Red Blood Cell Count Using the Sysmex XE-2100 Automated Hematology Analyzer
下载PDF
导出
摘要 目的探讨Sysmex XE-2100全自动血细胞分析仪计数有核红细胞的准确性。方法应用Sysmex XE-2100全自动血细胞分析仪筛查有核红细胞阳性的69例标本,用显微镜计数其外周血有核红细胞数量,比较血细胞计数及白细胞分类情况。结果 Sysmex XE-2100与显微镜计数法计数外周血有核红细胞数量有极好的相关性(r=0.992),2种方法计数结果的差异无统计学意义(P>0.05)。采用有核红细胞计数法,白细胞数较去除有核红细胞前减低(t检验,P<0.05),且淋巴细胞比例明显减低,而红细胞、血红蛋白及血小板均无明显变化(P>0.05)。结论 Sysmex XE-2100全自动血细胞分析仪可灵敏、准确、快速地自动计数外周血有核红细胞,有效避免有核红细胞对白细胞总数及分类的干扰。 Objective To evaluate the accuracy of Sysmex XE-2100 automated hematology analyzer on counting nucleated red blood cell(NRBC).Methods Nucleated red blood cells were counted from 69 positive specimens by Sysmex XE-2100automatedhematologyanalyzerandcomparedwiththemanualmethodwithmicroscopy.Results Therewasa good correlation between the manual and automatic method(r=0.992).No significant difference was observed between the counting results of two methods(P0.05).When the peripheral NRBC appeared,the counting of WBC was significantly higher than the value obtained when the NRBC was depleted and the proportion of lymphocytes was significantly lower(P0.05).While there was no difference among the red blood cells,hemoglobin and platelets(P0.05).Conclusion The NRBC counting by the XE-2100 automated hematology analyzer is sensitive,quick,accurate and reliable.
出处 《临床输血与检验》 CAS 2010年第3期221-224,共4页 Journal of Clinical Transfusion and Laboratory Medicine
关键词 XE-2100血细胞分析仪 有核红细胞 准确性 Sysmex XE-2100 automated hematology analyzer Nucleated red blood cell Accuracy
  • 相关文献

参考文献16

二级参考文献38

  • 1何晓莲,曲斌,朱以勇,赵军.小儿血象有核红细胞临床分析[J].黑龙江医学,1999(7):33-33. 被引量:3
  • 2束国防,高茂馗,芦慧霞.XE-2100血液分析仪对异常细胞报警的评估[J].东南大学学报(医学版),2005,24(2):98-100. 被引量:10
  • 3李建英.血细胞分析仪自动计数外周血有核红细胞的应用研究[J].中华检验医学杂志,2005,28(6):606-609. 被引量:23
  • 4无.全国血液学复检专家小组工作会议纪要暨血细胞自动计数复检标准释义[J].中华检验医学杂志,2007,30(4):380-382. 被引量:160
  • 5叶应妩 王毓三.全国临床检验操作规程[M].南京:东南大学出版社,1991.70-73.
  • 6BARNES PW, MCFADDEN SL, MACHIN SJ, et al. The international consensus group for hematology review :suggested criteria for action following automated CBC and WBC'differential analysis[J]. Lab He-matal, ,2005,11:83--90.
  • 7STACHON A, HOLLAND-LETZ T, KR1EG M. High in-hospital mortality of intensive care patients with nucleated red blood cells in blood[J]. Clin Chem Lab Med,2004,42 .. 933-- 938.
  • 8HUANG R,BARBER TA,SCHMIDT M A,et al. A microfluidics approach for the isolation of nucleated red blood cells (NRBCs) from the peripheral blood of pregnant women [J]. Prenat Diagn, 2008, 28 : 892 -- 899.
  • 9STEURER M A, BERGER T M. Massively elevated nucleated red blood cells and cerebral or pulmonary hemorrhage in severely growth--restricted infants- - is there more than coincidence[J]? Gerontology, 2008,94:314--319.
  • 10[4]Igout J,Fretigny M,Vasse M,et al.Evaluation of the CoIllter LH750 haematology analyzer compared witll flow cytometry as the reference method for WBC,platelet and nucleated RBC count[J].Clin Lab Hamatol,2004,26(1):1-7.

共引文献71