期刊文献+

Effect of HTS Superconductors on Homogeneity of Measurement Field in Low Field Nuclear Magnetic Resonance Detection

Effect of HTS Superconductors on Homogeneity of Measurement Field in Low Field Nuclear Magnetic Resonance Detection
下载PDF
导出
摘要 In recent years, superconducting quantum interference devices (SQUIDs) have been demonstrated to be useful in the low field nuclear magnetic resonance (NMR) measurements. The high temperature superconducting (HTS) SQUID used in our experiments has a frequency-independent sensitivity of 40-50fT/Hz^1/2. When a liquid nitrogen cooled LC circuit is employed to form a tuned circuit with the SQUID, the sensitivity of the system can be further enhanced. The LC circuit consists of a capacitor and a coil made of copper wire or HTS tape, which is inductively coupled to the SQUID. However, the homogeneity of the measurement field deteriorates because of the HTS tape coil in the proximity of the sample. In contrast, the thin film SQUID with a washer area of 1 cm^2 has no effect on the NMR signal. Therefore, the impairment of the measurement field homogeneity in the case of different superconducting elements nearby is discussed by examining the free induction decay signals at 9 kHz. It is found that a square superconducting film with an area of i cm^2 may compensate for the inhomogeneity of the measurement field after the adjustment of its position. In recent years, superconducting quantum interference devices (SQUIDs) have been demonstrated to be useful in the low field nuclear magnetic resonance (NMR) measurements. The high temperature superconducting (HTS) SQUID used in our experiments has a frequency-independent sensitivity of 40-50fT/Hz^1/2. When a liquid nitrogen cooled LC circuit is employed to form a tuned circuit with the SQUID, the sensitivity of the system can be further enhanced. The LC circuit consists of a capacitor and a coil made of copper wire or HTS tape, which is inductively coupled to the SQUID. However, the homogeneity of the measurement field deteriorates because of the HTS tape coil in the proximity of the sample. In contrast, the thin film SQUID with a washer area of 1 cm^2 has no effect on the NMR signal. Therefore, the impairment of the measurement field homogeneity in the case of different superconducting elements nearby is discussed by examining the free induction decay signals at 9 kHz. It is found that a square superconducting film with an area of i cm^2 may compensate for the inhomogeneity of the measurement field after the adjustment of its position.
出处 《Chinese Physics Letters》 SCIE CAS CSCD 2010年第8期250-252,共3页 中国物理快报(英文版)
  • 相关文献

参考文献10

  • 1Packard M E and Varian R 1954 Phys. Rev. 93 941.
  • 2Clarke J, Lee A T, Muck M and Richards P L 2006 The SQUID Handbook ed Clarke J and Braginski A I (Weinheim: Wiley-VCH) vol Ⅱ and references therein.
  • 3Bernarding J, Buntkowsky G, Macholl S, Hartwig S, Burghoff M and Trahms L 2006 J. Am. Chem. Soc. 128 714.
  • 4Qiu L, Zhang Y, Krause H -J, Braginski A I, Burghoff M and Trahms L 2007 Appl. Phys. Lett. 91 072505.
  • 5Qiu L, Zhang Y, Krause H -J, Braginski A I and Usoskin A 2007 Rev. Sci. Instrum. 78 054701.
  • 6Dong H, Zhang Y, Krause H -J, Xie X, Braginski A I and Offenhausser A 2010 J. Conf.: Phys. Series (accepted).
  • 7Zhang Y, Qiu L, Krause H -J, Dong H, Braginski A I and Offenhausser A 2009 Physica C 469 1624.
  • 8Dong H, Zhang Y, Krause H -J, Xie X, Braginski A I and Offenhausser A 2009 Supercond. Sci. Technol. 22 125022.
  • 9The superconducting coil was fabricated by European High Temperature Superconducors GmbH & Co. KG (EHTS).
  • 10Zhang Y, Schubert J and Wolters N 2002 Physica C 372- 376 282.

相关作者