摘要 提出左心机械辅助的一种新设想动力性主动脉瓣,在主动脉瓣的位置设置一可主动旋转的螺旋桨式机械性瓣叶(叶轮)。瓣叶由置于升主动脉腔内的支架笼中心的一对轴承支承。支架笼通过其外周的编织瓣裙与主动脉瓣环缝合固定。笼内有可自由旋转的转子体,笼架支承主动脉壁从而防止转子与动脉壁接触。转子体与旋转瓣叶通过一共同轴联接并传递动力。将筒状的定子置于主动脉壁外,其功能与普通的电机相同。当接通电源时,定子产生的磁场将驱动叶轮转子体旋转推动血液由心室进入主动脉,达到心室卸负荷的目的。本文对该设想进行了模型研究,结果表明原理正确可行,有必要进一步探索。 Dynamic aortic valve, a novel design for left ventricular assistance is proposed. An impeller similar to the airscrew is placed in the position of aortic valve. The impeller is held by a pair of bearings in the center of a cylindrical supporting cage which is located in the lumen of the aorta ascendens. The supporting cage is attached to the valve annulus by a sewing ring of textile cuff around the outside of the cage. A rotor is fitted into the cage and connected to the impeller with a common shaft. The cage supports the aortic wall,thus prevents the rotor from fouching the wall and allows the rotor to rotate freely. A stator, functioning as that of the electric motor, is located outside the aortic wall. When actuated, the magnetic field generated by the stator would drive the rotor to turning, and the axial flow of blood from ventricular to the aorta would be produced, resulting in the reduction of the ventricular load. We studied this concept in a simulating model. The result suggests that this design be reasonable. Further exploration of this concept is justified.
出处 《生物医学工程学杂志》 EI CAS CSCD 北大核心 1999年第1期116-119,共4页 Journal of Biomedical Engineering
关键词 动力性 主动脉瓣 左心辅助 旋转叶轮泵 Dynamic aortic valve Left ventricular assistance Rotary pump
  • 相关文献

参考文献1

  • 1Levin H R,Ann Thorac Surg,1994年,58卷,1515页

同被引文献89

  • 1曹小妹,凌晓.医用高分子材料的抗凝血性研究进展[J].湖南师范大学学报(医学版),2000(4):74-75. 被引量:2
  • 2钱坤喜,王颢,茹伟民,曾培,袁海宇.轴向驱动永磁磁浮离心血泵的试制型装置[J].江苏大学学报(自然科学版),2005,26(6):510-513. 被引量:7
  • 3宋述军,彭燕,沙次文,凌金福.交流磁流体推进血液驱动装置的初步研究[J].中国生物医学工程学报,2006,25(5):557-562. 被引量:5
  • 4徐先懂,谭建平.基于磁能传递的血泵驱动系统研究[J].机械,2006,33(10):7-9. 被引量:3
  • 5Leslie W. Patient Slection for the Use of Ventricular Assist Devicees as a Bridge to Transplantation. Ann Thorac Surg,2003, 75 :S67 -71.
  • 6Akutsu T,Koyanagi H. ARTIFICIAL HEART 3. Springer-Verlag Tokyo. Pringted in Hong Kong,1991:6.
  • 7Loebe M, Gorman K, Burger R. Complement activation in patients undergoing mechanical circulatory support. ASAIO J, 1998, ( 5 ) : M340 - 6.
  • 8Corry DC,DeLucia A, Zhu H. Time course of cytokine release and complement activation after implantation of the HeartMate left ventricular assist device. ASAIO J, 1998,44 (5) : M347 - 51.
  • 9Loebe M, Koster A, Sanger S. lnnammatory response after implantation of a left ventricular assist device: comparison between the axial flow MicroMed DeBakey VAD and the pulsatile Novacor device. ASAIO J, 2001,47(3 ) :272 -4.
  • 10Spanier TB, Oz M, Rose E, et al. Activation of NF-kappa B is central to the proinflammatory/procoagulant response in textured surface left ventricular assist device recipients and may be influenced by anti-inflammatory intervention with aspirn. J Heart Lung Transplant, Abstract Issue, 18th Annual Meeting and Scientific Sessions, Chicago, IL, 1998.