期刊文献+

用无二氧化碳水代替去离子除氧水测定硫化物 被引量:1

Determination of Sulfide with None Carbon Dioxide Contained Water Replacing Deionizing and Deoxidizing Water
下载PDF
导出
摘要 在新颁国标中,采用亚甲基蓝光度法测定水和废水中硫化物时,要求用去离子除氧水做为实验用水。在日常工作中,用无二氧化碳水代替去离子除氧水,二者的空白吸光度和标准曲线无显著性差异,且其精密度均小于5%,准确度在93%-103%之间。 The deionizing and deoxidizing water is required by the methylene blue spectrophotometric method in the new national standard to analyze sulfide in water and waste water. The non-carbon dioxide contained water and deionizing and deoxidizing water are used in the comparison experiment of sulfile measurement.The blank photometric absorption and the standard cure of them aren't remarkable discrepancy. The precision of them are all less than 5%, and the accuracy are in 93%-103%.
作者 李静
出处 《石油与天然气化工》 CAS CSCD 北大核心 1999年第1期68-70,共3页 Chemical engineering of oil & gas
关键词 亚甲基蓝光度法 硫化物 水质硫化物 水质监测 methylene blue spectrophotometric method, sulfide, deionizing and deoxidizing water, noncarbon dioxide contained water
  • 相关文献

参考文献4

  • 1国家环保局.水和废水监测分析方法(第三版)[M].北京:中国环境科学出版社,1989.108,354.
  • 2国家环保局.水质硫化物的测定(亚甲基蓝分光光度法)[M].北京:中国环境科学出版社,1996..
  • 3国家环保局,水质硫化物的测定(亚甲基蓝分光光度法),GB/T 16489-1996,1996年
  • 4国家环保局,水和废水监测分析方法(第3版),1989年,326页