椎弓根螺钉长度变化对螺钉-骨复合体模型应力影响的三维有限元分析研究 被引量:18

Three dimensional finite element analysis of stress distribution on continuously varying of length of pedicle screw
下载PDF
导出
摘要 目的利用三维有限元模型研究椎弓根螺钉长度变化对生理载荷下螺钉-骨复合体模型的应用影响。方法建立椎弓根螺钉和L1椎体的三维模型,并对其进行网格划分,设置椎弓根钉长度尺寸的变化范围。模拟生理载荷条件下,对不同长度尺寸的椎弓根钉有限元模型进行应力分析。结果椎弓根螺钉长度在30~50mm范围内变化时,随着螺钉长度的增大,螺钉-骨复合体模型的骨质部分承担的应力均减小,而螺钉承担的应力则增加。螺钉最大平均主应变出现在螺钉的尾端,皮质骨发生的最大平均主应力位置出现在螺钉与皮质骨接触面两侧,松质骨发生的最大平均主应力位置出现在螺钉头部与松质骨接触面两侧。当螺钉长度达到50mm时,载荷力传递到皮质骨和松质骨分别减小了43.1%和42.3%,而螺钉上出现的则增加了38%。当椎弓根螺钉长度大于45mm时,螺钉-骨复合体模型各部分应力变化不明显。结论椎弓根螺钉长度在30~50mm范围变化时,在生理载荷下,椎弓根螺钉长度的增大有利于改善螺钉、皮质骨及松质骨上轴向应力的力学分布;只要骨量允许,临床选择椎弓根螺钉的长度应不小于45mm。 Objective To study the stress distribution on different length of pedicle screws under the physiological load by using the three dimensional finite element analysis.Method A three-dimension model of the pedicle screw and L1 vertebral body were constructed with the model meshed on the basis of the finite element method.The feature dimension of pedicle screw in the model was set in a specified varied range.Under the physiological load,the stress on every model with different diameter sizes of pedicle screw was analyzed.Results The stress on every bone model decreased with the increase of stress on screw under the axially pullout force as the length of screw ranged from 30mm to 50mm.And the maximum Equivalent Stress(EQV Stress) in the pedicle screw appeared in the central area of the pedicle screw,the maximum Equivalent Stress(EQV Stress) in the cortical bone appeared in both sides of contact surface,the maximum Equivalent Stress(EQV Stress) of the cancellous bone appeared in both sides of contact surface of the top of pedicle screw and cancellous bone.In a certain external load,with the 50mm length of screw,the load that transfers to the cortical bone and cancellous bone is reduced by 43.1% and 42.3%,respectively,while the maximum Equivalent Stress(EQV Stress) of screw was increased 38%.When L≥45mm,the variable stress on all models become stable.Conclusions While the length of screw is in range of 4.0 mm to 6.5 mm,30~50 mm,the increase of pedicle screw length could improve the distribution of axial pullout stress on the screws,cortical bone and cancellous bone.As long as the bone mass allowed,the length of pedicle screws should be not less than 45mm in clinical choice.
出处 《医用生物力学》 EI CAS CSCD 2010年第3期206-211,共6页 Journal of Medical Biomechanics
基金 国家高技术研究(863)发展计划(2007AA02Z468)
关键词 螺钉 有限元分析 抗拔出力 平均主应力 应力 Screw Finite element analysis Pullout force Equivalent stress Stress
  • 相关文献

参考文献19

  • 1Hajek PD, Lipka J, HarUine P, et al. Biomechanical study of C1-C2 posterior arthrodesis techniques [ J ]. Spine,1993,18(2) :173-177.
  • 2Chen CS, Chen WJ, Cheng CK, et al. Failure analysis of broken pedicle screws on spinal instrumentation [ J ]. Med Eng P hys, 2005, 27 (6) :487-496.
  • 3McKinley TO, McLain RF, Yerby SA, et al. Characteristics of pedicle screw loading. Effect of surgical technique on intravertebral and intrapedicular bending moments [ J ]. Spine, 1999,24( 1 ) :18-24.
  • 4Misenhimer GR, Peek RD, Wiltse LL, et al. Anatomic analysis of pedicle cortical and cancellous diameter as related to screw size[J]. Spine, 1989,14(4) :367-372.
  • 5Panjabi MM, Shin EK, Chen NC, et al. Internal morphology of human cervical pedicles [ J ]. Spine, 2000,25 (10) : 1197-1205.
  • 6Walker MD, McMahon DJ, Udesky J, et aL Application of high-resolution skeletal imaging to measurements of volumetric BMD and skeletal microarchitecture in Chinese- American and white women:explanation of a paradox[ J ]. J Bone Miner Res, 2009,24(12) :1953-1959.
  • 7El-Rich M, Arnoux PJ, Wagnac E, et aL Finite element investigation of the loading rate effect on the spinal load-sharing changes under impact conditions [J]. J Biomech, 2009,42(9) : 1252-1262.
  • 8Hashemi A, Bednar D, Ziada S. Pullout strength of pedicle screws augmented with particulate calcium phosphate: an experimental study[ J ]. Spine J, 2009,9 (5) :404-4]0.
  • 9牛文鑫,丁祖泉.三种三维有限元建模方法在跟骨模型建立中的应用和比较[J].医用生物力学,2007,22(4):345-350. 被引量:20
  • 10Lafage V, Gangnet N, Senegas J, et al. New interspinous implant evaluation using an in vitro biomechanical study combined with a finite-element analysis [ J]. Spine, 2007,32 (16) : 1706-1713.

二级参考文献45

  • 1杨国标,徐峰.基于Hangmann骨折的植骨融合钢板内固定术有限元稳定性分析[J].力学季刊,2005,26(3):491-496. 被引量:4
  • 2佟大可,蔡郑东,吴建国,许硕贵,纪方.骶骨肿瘤切除术后骨盆三维有限元模型的建立及相关应力分析[J].医用生物力学,2005,20(4):231-234. 被引量:11
  • 3Belytschko T, Kulak RF, Schultz AB, et al. Finite element stress analysis of an intervertebral disc [J]. J Biomech, 1974,7(3):277-285.
  • 4Liu YK, Ray G, Hirsch. The resistance of the lumber spine to direct shear [J]. Orthop Clin North Am, 1975,6(1):33-49.
  • 5Yoganandan N, Kumaresan S, Voo L, et al. Finite element applications in human cervical spine modeling [J]. Spine, 1996,21 (15):1824-1834.
  • 6Puttlitz CM, Goel VK, Clark CR, et al. Pathomechanisms of failure of the odontoid [J]. Spine, 2000,25(22):2868-2876.
  • 7Graham RS, Oberlander EK, Stewart JE, et al. Validation and use of a finite element model of C-2 for determination of stress and fracture patterns of anterior odontoid loads [J]. J Neurosurg, 2000,93(Suppl 1):117-125.
  • 8Teo EC, Ng HW. First cervical vertebra (atlas) fracture mechanism studies using finite element method [J]. J Biomech, 2001,34(1):13-21.
  • 9Bozkus H, Karakas A, Hanci M, et al. Finite element model of the Jefferson fracture: comparison with a cadaver model [J]. Eur Spine J, 2001,10(3):257-263.
  • 10Stehlik J, Bartos M, Kestranek Z, et al. Application of numerical modelling of osteotomy to orthopaedic practice [J]. Int J Med Inf,1997,45(1-2):75-82.

共引文献50

同被引文献209

引证文献18

二级引证文献113