摘要
建立HPLC—MS/MS联用方法定量测定人血浆中恩替卡韦浓度。以地西泮为内标,甲醇为有机相,甲醇一水为(5mmol·L^-1碳酸氢铵)梯度洗脱流动相,Waters—Eterra MS-C18(2.1×50mm×5pm)色谱柱为分析柱,通过电喷雾离子源(ESI),以正离子多反应监测(MRM)方式检测。用于定量分析的离子对分别为,n/z278.2—152.1(恩替卡韦)和m/z285.0→193.0(地西泮,内标)。恩替卡韦的线性范围为0.05~20μg·L^-1,定量下限为0.05μg·L^-1(n=6)。日内、日问精密度的RSD〈15%,平均回收率〉75%。
Entecavir in human plasma was determined by high performance liquid coupled with-tandem mass spectrometry (HPLC-MS/MS). Diazepam was used as internal standard. Entecavir was separated on a Waters-Eterra MS-C18 column(2.1 ×50 mm× 5 μm). Electro- spray ionization(ESI) source was applied, and multiple reaction monitoring(MRM) mode was operated in the positive mode with the monitor ions at m/z 278.2→152.1 for entecavir and m/z 285.0→193.0 for the internal standard, respectively. The linear calibration curve is obtained over the concentration range of 0.05-20 μg · L^- 1. The limit of quantitation is 0.05 μg · L^-1 (n=6). The inter and intra-day precision(RSD) is less than 15%. The average recoveries are above 75%.
出处
《质谱学报》
EI
CAS
CSCD
2010年第4期252-256,共5页
Journal of Chinese Mass Spectrometry Society
关键词
恩替卡韦
高效液湘色谱-串联质谱
血浆药物浓度
entecavir
high performance liquid chromatography coupled with-tandem mass spectrometry (HPLC-MS/MS)
plasma concentration