提高真空灭菌器B-D测试的效果监测研究 被引量:1

Study of improving monitoring effect of B-D test on vacuum sterilizer
下载PDF
导出
摘要 目的提高消毒人员对B-D测试假阳性结果判断的准确率,提高真空灭菌器的正常使用率。方法将B-D测试条、1292嗜热脂肪杆菌芽孢、132℃包内化学指示卡同时放入自制B-D包中央作为对照组,另将相同B-D测试条放入螺旋管内作为试验组,共试验92锅次。观察2组灭菌监测结果。结果试验组阴性率为100.00%高于对照组的81.52%,差异有统计学意义(P<0.05)。结论在规范操作下用螺旋管放置B-D测试条可以明显提高该化学指示条的变色合格率,提高了消毒人员对B-D测试的假阳性结果判断的准确率,从而保证了灭菌的安全性。 Objective To increase the judgment's accuracy rate to false positive results of B-D test by disinfecting personnel,and improve the normal usage of vacuum sterilizer.Methods Put B-D testing strips,1292 thermophilic fat bacilli germ and in-bags' chemical instructions of 132℃ in the central of self-made B-D bags as the control group;Put the same B-D testing strips into canals of corti as the test group.Tested 92 pot times in all.Observed the monitering effects of sterilization.Results The negative rate of the test group was 100.00% higher than 81.52% of the control group,and the difference was statistically significant(P 0.05).Conclusion Put the B-D testing strips into canals of Corti under the standard operation can significantly improve the qualified colour changed rate of chemical instructions,which avoid disinfecting personnel getting false positive results of B-D test,so as to ensure the safety of sterilization.
作者 黎平
出处 《临床合理用药杂志》 2010年第15期34-35,共2页 Chinese Journal of Clinical Rational Drug Use
关键词 B-D试验 螺旋管装置 真空灭菌器 效果监测 B-D test apparatus of canals of corti Vacuum sterilizer Monitoring effect
  • 引文网络
  • 相关文献

参考文献6

二级参考文献18

  • 1符爱娟,陈桂.B-D试验不合格原因分析[J].中国消毒学杂志,2005,22(1):112-113. 被引量:4
  • 2李斌,魏静蓉,王义辉,唐伟.脉动真空灭菌器B-D试验阳性结果的分析与验证[J].医疗卫生装备,2007,28(3):61-62. 被引量:3
  • 3.消毒技术规范[S].[S].中华人民共和国卫生部,2002,11..
  • 4叶秀菊,叶晓玲,高丽萍.B-D试验测试不合格原因排查[J].中国消毒学杂志,2007,24(5):483-484. 被引量:3
  • 5李新武.国外医疗机构灭菌保障体系初步探讨.中华医院感染学杂志,2008,18(8):1118-1118.
  • 6International Organization of Standardization. ISO11140-4,Sterilization of health care products-chemical indicators-part4:Class 2 indicators as an altemative to the Bowie and Dick-type test for detection of steam penetration [ S ]. 2007 : 30.
  • 7International Organization of Standardization. ISO11140-5,Sterilization of health care products-Chemical indicators-partS:Class 2 indicators for Bowie and Dick-type air removal tests[S]. 2007:9.
  • 8European Norm. EN285:2006 Steam Sterilizer-Large Sterilizer[S]. European Committee for Standardization,2006:44.
  • 9Association for the Advancement of AMMI ST 79: Comprehensive guide assurance in the health care facilities 2006:10 Medical Instrumentation, ANSI/ to steam sterilization and sterility [S]. American National Stand,
  • 10任伍爱.B-D测试新进展及应用[J].中华医院感染学杂志,2008,18(9):1349-1350. 被引量:7

共引文献1245

同被引文献6

引证文献1

;
使用帮助 返回顶部