PAGAT型聚合物凝胶剂量计的高熔点优化 被引量:1

An Improvement for Polymer Gel Dosimeter of Type PAGAT
下载PDF
导出
摘要 使用丙烯酰胺、明胶、交联剂、除氧剂、甲醛和琼脂糖等,在常氧条件下制备出四种不同配比的聚合物凝胶剂量计。根据U型管两端水平液面恒等原理,在32℃和60℃下比对筛选出具有较高熔点的凝胶剂量计,实验结果表明PAGAT加入适量甲醛所得的聚合物凝胶剂量计具有60℃以上的熔点,可以很好抵御环境温度的大幅度变化,并且有很好的剂量线性响应特性。 Four types of polymer gel dosimeter, containing aerylamide, gelatin, BIS, THPC, formaldehyde and agarose, were prepared under normal atmospheric conditions. Aecording to the principle of Utype test tube, we screened out the gel with high melting point at 32℃ and 60℃. It turned out that the PAGAT gel mixed with formaldehyde had a melting point over 60℃, which made the gel resist the varied environment temperature to a great extent. And it also had a good liner response with its absorbed dose.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2010年第7期935-939,共5页 Nuclear Electronics & Detection Technology
基金 国家自然科学基金资助项目(10775178)
关键词 聚合物凝胶剂量计PAGAT熔点 polymer gel dosimeter, PAGAT, melting point
  • 相关文献

参考文献17

  • 1Geoffrey S.Ibbott.Clinical Applications of Gel Dosimeters[J].Journal of Physics:Conference Series,2006,56:108-131.
  • 2Clive Baldock,Historical overview of gel dosimetry[J].Journal of Physics:Conference Series,2004,3:1-3.
  • 3SvenJ B ck.Clinical applications of gel dosimeters[J].Journal of Physics:Conference Series,2009,164:012049.
  • 4Clive Baldock.Historical overview of the development of gel dosimetry:another personal perspective[J].Journal of Physics:Conference series,2009,164:012002.
  • 5Keall P J and Baldock C.A theoretical study of the radiological properties and water equivalence of Fricke and polymer gels used for radiation dosimetry Australas[J].Phys.Eng.Sci.Med,1999,22:85 91.
  • 6Gore J,Kang Y and Schulz R.Measurement of radiation dose distributions by nuclear magnetic resonance (NMR) imaging[J].Phys Med Biol,1984,29:1189-1197.
  • 7Olsson L,Westrin B,Fransson A,et al.Diffusion of ferrous inos in agarose dosimeter gels[J].Phys Med Biol,1992,37:2243-2252.
  • 8Baldock C,Harris P J,Piercy A R,et al.Experimental determination of the diffusion coefficient in two-dimensions in ferrous sulphate gels using the finite element method Australas[J].Phys.Eng.Sci.Med,2001,24:19-30.
  • 9Maryanski M J,Gore J C,Kennan R P,et al.NMR relaxation enhancement in gels polymerized and cross-linked by ionizing radiation:a new approach to 3D dosimetry by MRI Magn.Reson.Imaging,1993,11:253-258.
  • 10Maryanski M J,Schulz R J,Ibbott G S,et al.Magnetic resonance imaging of radiation dose distributions using a polymer-gel dosimeter[J].Phys.Med.Biol,1994,39:1437-1455.

同被引文献14

  • 1何承发,巴维真,吾勤之,王倩.剂量增强效应的半经验理论计算[J].核技术,1997,20(2):95-99. 被引量:2
  • 2IBBOTT G S. Clinical applications of gel dosimeters[J]. J Phys, 2006, 56: 108-131.
  • 3BALDOCK C. Historical overview of gel dosimetry[J]. J Phys, 2004, 3: 1-3.
  • 4MAYANSKIT M J. Magnetic resonance imaging of radiation dose distributions using a polymergel dosimeter [J]. Phys Med Biol, 1994, 39: 1 437-1 455.
  • 5HURLEY C. A study of a normoxic polymer gel dosimeter comprising methacrylic acid, gelatin and tetrakis (hydroxymethyl) phosphonium chloride (MAGAT)[J]. Appl Radiat Isot, 2005, 63: 443-456.
  • 6AJBACK S. Clinical applications of gel dosimeters[J]. J Phys, 2009, 164: 012049.
  • 7BALDOCK C. Historical overview of the development of gel dosimetry: Another personal perspeetive[J]. J Phys, 2009, 164: 012002.
  • 8SCHREINER L J. NMR mechanisms in gel dosimetry[J]. J Phys, 2009, 164:012032.
  • 9HILL B. The dose response of normoxic polymer gel dosimeters measured using X-ray CT[J]. The British Journal of Radiology, 2005, 78 : 623-630.
  • 10POURFALLAH T A, ALLAHVERDI M,ALAM N R, et al. Performance evaluation of MRI-based PAGAT polymer gel dosimeter in an inhomogeneous phantom using EGSnrc code on a Co-60 machine[J]. Appl Radiat Isot, 2009, 67: 186-191.

引证文献1