金属氧化物半导体场效应晶体管测量组织补偿胶对乳腺癌胸壁照射皮肤X线表面剂量的影响

Investigation about the effect of tissue compensation glue on surface dosage in chest wall radiation of breast cancer by using MOSFET detector
下载PDF
导出
摘要 目的使用金属氧化物半导体场效应晶体管(metal oxide semiconductor field effect transistor,MOSFET)探测器测量不同厚度组织补偿胶对乳腺癌胸壁照射X线皮肤表面剂量的影响。方法选用6MV、15MV X线进行照射,分为有补偿胶组和无补偿胶组,补偿胶按厚度分为3mm、5mm、10mm、15mm、20mm5组,用MOSFET分别测量,每组测10次。结果使用组织补偿胶后表面剂量明显增加,不同厚度补偿胶对表面剂量的影响程度不同。结论使用补偿胶后射线的剂量建成区出现变化,表面剂量明显提高。临床治疗中需要根据肿瘤类型及深度选用合适的补偿胶。 Objective Using MOSFET detector to measure the effect of tissue compensation glue on surface dosage in chest wall radiation of breast cancer by using MOSFET detector.Methods Expose the experimental objects to 6MV,15MV X-rays and the objects have already been divided into glued group and non-glued group before the exposure.Using MOSFET to measure the dose of each group for 10 times and 5 types of tissue compensation glue of different thickness were measured.Results The surface dose increased significantly after applying the tissue compensation glue;and different types of glues had their identical effect on it.Conclusion The build-up regions have changed and surface dosage increases considerably after the application of tissue compensation glue.Hence,proper attentions should be focus on the selection of proper glue in chest wall radiation of breast cancer.
出处 《中国现代医药杂志》 2010年第8期9-11,共3页 Modern Medicine Journal of China
关键词 MOSFET 组织补偿 表面剂量 乳腺癌 MOSFET Tissue compensation glue Surface dosage Breast cancer
  • 相关文献

参考文献7

  • 1钟鹤立,顾伟英,张光伟,刘剑锋,李隆兴,吴何苟,陆佩芳.乳腺癌根治术后胸壁放疗技术与剂量分布[J].中国医学物理学杂志,2000,17(3):136-137. 被引量:9
  • 2惠周光,余子豪.乳腺癌的放射治疗[J].实用肿瘤杂志,2002,17(6):370-372. 被引量:18
  • 3Scalchi P, Francescon P. Calibration of a mosfet detection system for 6-MV in vivo dosimetry. Radiat Oncol Biol Phys,1998,40(4): 987-993.
  • 4Wu A,Maitz A,Shortt K,et al. Performance of a new dosimetry system with MOSFET sensor for radiosurgery applications. AAPM, Anaheim,CA, 1994,7:25-28.
  • 5张绍刚,于世平,刘原照,程金生,李开宝.平行板电离室对全身皮肤电子束治疗的吸收剂量测量[J].中华放射医学与防护杂志,2000,20(1):60-63. 被引量:2
  • 6Jornet N,Carrasco P,Jurado D,et al. Comparison study of MOSFET detectors and diodes for entrance in vivo dosimetry in 18 MV x-ray beams. Med Phys,2004,31(9):2534-2542.
  • 7Ramani R,Russel S,O'Brien P. Clinical dosimetry using MOSFETs. Inte J Radiat Oncol Biol Phys,1997,37(4):959-964.

二级参考文献5

共引文献25

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部