Infusa Sleeve导管血管内局部给药的实验研究 被引量:3

Experimental Study on Intravascular Local Drug Delivery With Infusa Sleeve Catheter
下载PDF
导出
摘要 目的:评价InfusaSleeve(IS)导管局部给药的可行性。方法:13只家兔随机分成A组(n=8)和B组(n=5),给药压力分别为607.9kPa(6个标准大气压)(A组)和202.6kPa(2个标准大气压)(B组),用标准血管成形导管损伤腹主动脉下端后,经IS导管向腹主动脉下端球囊损伤处局部注入1ml放射性同位素锝标记的甲氧基异丁基异腈(99mTc-MIBI)注射液。术后30分钟、6小时、24小时分别处死动物后,取两组动物靶点血管(腹主动脉下端)、心、肝、骨骼肌、小肠、血液、非靶点血管(肺动脉)组织,用γ计数仪计数后称组织重量。结果:A组和B组给药后30分钟靶点血管局部的放射强度分别占给药总量的49.250%和22.000%、分别是血液的45倍和8倍、是非靶点血管的70倍和12倍。靶点血管局部放射强度A组大于B组(P<0.05)。6小时和24小时后靶点血管局部放射强度占给药总量的0.343%(A组)和0.255%(A组)。结论:IS导管可经血管局部靶向给药。IS导管局部导入的药物浓度呈压力依赖性。 Objective:To assess the feasibility of the Infusa Sleeve (IS) catheter for the delivery of an agent into arterial wall at the site of percutaneous transluminal coronary angioplasty (PTCA) injury. Methods:Thirteen rabbits were randomly divided into two groups (A and B).Following a lesion dilatation of the wall of abdominal aorta by a standard angioplastic catheter,1 ml of Technetium[ 99m Tc] methoxy isobutyl isonitrile injection ( 99m Tc MIBI) was locally delivered to the injured site with IS catheter,at 607 9 kPa(6 atm) (group A, n =8) and 202 6 kPa (2 atm) (group B, n =5) pressures,respectively.The rabbits were sacrificed at 30 minutes,6 hours and 24 hours after the drug delivery.The target artery,control artery (pulmonary artery),cardiac muscle,liver,intestine and skeletal muscle were surgically isolated and scintillation counting and weighing were performed. Results:Thirty minutes following 99m Tc MIBI delivery,the radioactivity(Te cpm)of the target vessel wall as a percentage of in vitro value was 49 250% in group A,and 22 000% in group B;It was about 45 fold and 8 fold higher than that of the blood and 70 fold and 12 fold higher than that of control artery in group A and group B,respectively.The difference between the two groups was signficant ( p <0 05).The other 2 time point results showed that the radioactivity remained detectable (0 255%) up to 24 hours. Conclusions:It is possible to deliver agent to the local vessel wall via IS catheter;Agent delivery via the IS is pressure dependent.Evidence of 99m Tc MIBI delivery persists at least 24 hours.
出处 《中国循环杂志》 CSCD 北大核心 1999年第2期81-83,共3页 Chinese Circulation Journal
关键词 血管成形术 术后 再狭窄 预防 局部给药 IS导管 Local delivery Infusa Sleeve catheter Restenosis
  • 相关文献

参考文献1

  • 1Hong M K,Current Problems in Cardiology,1997年,22卷,1页

同被引文献19

  • 1[1]Serruys PW,de Jaegere P,Kiemeneij F,et al.A comparison of balloon expandable stent implantation with balloon angioplasty in patients with coronary artery disease [J]. N Engl J Med ,1994,331:489-495.
  • 2[2]Fischman DL,Leon MB,Baim DS,et al.A randomized comparison of coronary stent placement and balloon angioplasty in the treatment of coronary artery disease [J]. N Engl J Med ,1994,331:496-501.
  • 3[3]Erbel R,Haude M,Hopp HW,et al.Coronary artery stenting compared with balloon angioplasty for restenosis after initial balloon angioplasty [J]. N Engl J Med, 1998, 339:1672-1678.
  • 4[4]Miano JM,Tota RR,Vlasic N,et al.Early proto-oncogene expression in rat smooth muscle cell following endothelial removal[J]. Am J Pathol ,1990, 137:761-765.
  • 5[5]Shi Y,Fard A,Galeo A,et al.Transcatheter delivery of c-myc antisense oligomers reduces neointimal formation in a porcine model of coronary artery balloon injury[J]. Circulation, 1994, 90:944-951.
  • 6[7]Bauriedel G,Schluckebier S,Hutter R,et al.Apoptosis in restenosis versus stable angina atherosclerosis implication for the pathogenesis of restenosis [J]. Arterioscler Thromb Vasc Biol, 1998, 18:1132-1139.
  • 7[8]Bennett MR, Evan GI,Newby AC, et al. Deregulated expression of the c-myc oncogene abolishes inhibition of proliferation of rat vascular smooth muscle cells by serum reduction;interferon-γ;heparin and cyclic nucleotide analogues and induces apoptosis [J]. Circ Res, 1994,74:525-536.
  • 8[10]Kollum M,Kaiser S,Kinscherf R,et al. Apoptosis after stent implantation compared with balloon angioplasty in rabbits :Role of macro-phages [J]. Arterioscler Thromb Vasc Biol, 1997,17:2383-2388.
  • 9Hong MK, Roxana, Mintz GS, et al. Restenosis after coronary angioplasty [J]. Curr Probl Cardiol, 1997,22(1) : 1 - 36.
  • 10Tomaru T,Fujimor Y,Morita T,et al.Local delivery of antithrombotic drug prevents restenosis after balloon angioplasty in atherosclerotic rabbit artery[J].Jpn Circ J,1996,60(3):981 —922.

引证文献3

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部