IOL-Master在高度近视白内障人工晶状体度数测量中的应用 被引量:2

Biometric measurement by IOL-Master in cataract patients with high myopia
原文传递
导出
摘要 目的 比较IOL-Master和A型超声测量高度轴性近视白内障患者眼轴的准确性. 方法分析2008年9月至2009年9月行白内障超声乳化摘除及人工晶状体植入术的高度轴性近视白内障患者18例(28只眼),分别用接触式A超、IOL-Master两种方法测量眼轴长度,IOL-Master测量角膜曲率,SRK/T公式计算人工晶体度数,术后3个月随诊检查视力和屈光度. 结果接触式A超测量的眼轴均值为(28.71±1.14)mm,IOL-Master为(29.31±1.27)mm;接触式A超和IOL-Master间差异有统计学意义(P <0.05).结论 IOL-Master测量眼轴准确性较高;而A超方法测量变异度较大;IOL-Master准确性、非接触性对高度轴性近视白内障手术人工晶状体度数的选择有重要意义. Objective To compare the measurement of axial length (AL) by IOL-Master and contact ultrasonic (US) axial scan (A-scan) in high myopia with cataract. Methods The data of 18 patients (28 eyes) who were all high myopia with cataract and accepted the phacoemulsification and foldable IOL implantation in our hospital from September in 2008 to September in 2009 were analyze.The measurement of axial length was performed in 28 eyes by contact A-scan and IOL-Master, respectively. Keratometric power was measured by IOL-Master. And IOL power calculation was carried out according to SRK/T formula in the basis of IOL-Master. The cases were followed up for one month. Results The axial length was (28.71±1.14)mm by contact A -scan, (29.31±1.27)mm IOL-Master, respectively. There were significant difference between contact A-scan and IOL-Master (P 〈0.05). Conclusions The IOL-Master is more accurate than the contact A-scan, and IOL-Master is an better choice to measure the axial length in high myopia with cataract.
出处 《中国实用眼科杂志》 CSCD 北大核心 2010年第8期874-876,共3页 Chinese Journal of Practical Ophthalmology
关键词 眼轴长度 IOL-MASTER 接触式A超 近视 Axial length IOL-Master Ultrasonic scanning cataract Myopia
  • 相关文献

参考文献9

  • 1Wunder H.Increase A-scan accuracy for improved outcomes[J].Review of ophthalmology,2003,10(1):36-38.
  • 2H.John Shammas.1ntraocular Lens Power Calculation[M].America:SLACK lncoporated,2004:161.
  • 3汤萍,潘永称.高度近视白内障患者人工晶状体屈光度数计算公式的选择[J].中华眼科杂志,2003,39(5):290-293. 被引量:58
  • 4董晓光,谢立信,张怡,侯玲梅,纪惠谦,杨雪萍.人工晶体植入治疗高度近视并发白内障的临床观察[J].中华眼科杂志,1995,31(4):268-270. 被引量:19
  • 5戴锦晖,褚仁远,陆国生.高度轴性近视眼白内障手术屈光误差分析[J].中国实用眼科杂志,1999,17(10):595-597. 被引量:40
  • 6Packer M,Fine H,Hoffman RS.Immersion A-Scan compared with partial coherence interferometry outcomes analysis[J].Cataract Refract Surg,2002,28:239-242.
  • 7R.Bruce WallaceⅢ.Refractive cataract surgery and multifocal iols[J].Canada:SLACK lncoporated,2001,11:2-31.
  • 8Binkhorst RD.The accuracy of ultrasonic measurement of the axial eye length[J].Ophthalmic Surg,1981,12(5):363-365.
  • 9Mana Tehrani,Frank Krummenauer,Rajiv Kurnar,et a1.Comparison of biometric measurements using partial coherence interferometry and applanation ultrasound[J].J Cataract Refract Surg,2003,29(4):747-75l.

二级参考文献12

  • 1陆国生,中华眼科杂志,1993年,29卷,16页
  • 2李美玉,中华眼科杂志,1990年,26卷,131页
  • 3杨文辉,中华眼科杂志,1990年,26卷,337页
  • 4Seitz B;Langenbucher A;Nguyen N.Underestimation of intraocular lens power for cataract surgery after myopic photorefractive keratectomy[J],1999(4).
  • 5Lyle WA;Jin GJ.Phacoemulsification with intraocular lens implantation in high myopia,1996.
  • 6Jimenez Alfaro I;Miguelez S;Bueno JL.Clear lens extraction and implantation of negative-power posterior chamber intraocular lenses to correct extreme myopia[J],1998.
  • 7Hoffer KJ.The Hoffer Q formula: a comparison of theoretic and regression formulas,1993.
  • 8Kora Y;Yaguchi S;Inatomi M.Preferred postoperative refraction after cataract surgery for high myopia,1995.
  • 9Bardocci A;Lofoco G.Corneal topography and postoperative refraction after cataract phacoemulsification following radial keratotomy,1999.
  • 10王静,陈薇,张德荣,李建平,董秀芹.高度近视白内障超声乳化摘除及后房型人工晶体植入术[J].中华眼科杂志,1999,35(2):98-100. 被引量:50

共引文献103

同被引文献12

引证文献2

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部