摘要
在有一定高围压作用的动、静三轴冻土实验中,出现高压室内温度测量值不稳定以及突变现象,不是因为热敏电阻温度计本身受压后其特性变化所致,而是由于压力室内围压的变化(升高或降低)引起其温度场(油温)的变化所造成的.
in the dynamical or statical triaxialtests of frozen soils under a giver high confiningpressure, unstability or sudden change of observedtemperature occurs in the high presssure chamber.This phenomenon does not result from the changeo f thermistor itself under high prossure but fromthe change of temperature field in correspondencewith the confining pressure variation in the pressure chamber.
出处
《冰川冻土》
CSCD
北大核心
1999年第1期39-41,共3页
Journal of Glaciology and Geocryology
关键词
热敏电阻
温度计
围压
油温
温度计
冻土
: thermistor-thermometer, confining pressure, oil temperature