摘要
The article is the analysis of phenyl isocyanate in benzene by gas chromatography.Inspection column temperature、detector temperature and boil room temperature at the effect on the determination,determined the optimal conditions for gas chromatographic analysis:column temperature:180℃,detector temperature:250℃,boil room temperature:260 ℃.Under these conditions,the standard deviation are 0.10mg/L,the relative standard deviation were 0.48%,recoveries were more than 99.4%,achieved satisfactory results.Provides an analytic method of phenyl isocyanate in benzene.
The article is the analysis of phenyl isocyanate in benzene by gas chromatography.Inspection column temperature、detector temperature and boil room temperature at the effect on the determination,determined the optimal conditions for gas chromatographic analysis:column temperature:180℃,detector temperature:250℃,boil room temperature:260 ℃.Under these conditions,the standard deviation are 0.10mg/L,the relative standard deviation were 0.48%,recoveries were more than 99.4%,achieved satisfactory results.Provides an analytic method of phenyl isocyanate in benzene.
出处
《化学研究与应用》
CAS
CSCD
北大核心
2010年第9期1214-1216,共3页
Chemical Research and Application
基金
成都中医学大学校基金资助
关键词
异氰酸苯酯
气相色谱
苯
gas chromatography
phenyl isocyanate
Benzene