摘要 报道了EA1110型元素分析仪CHN模式下反应管填充方法的改进措施,燃烧管中按顺序填装银丝、Cr2O3、Co3O4+Ag;还原管只填装线状铜.同时针对不同样品采用不同的制样措施,保证了分析结果的准确性,燃烧管的平均使用次数延长了24%,还原管的平均使用次数延长了59%,提高了仪器使用效率. An improved method on filling reaction tubes in EA1110 element analyzer at CHN model is stated.In combustion tube,silver-wire,Cr2O3 and Co3O4+Ag are filled in order,and linear Cu is the unique reagent in reduction tube.Relevant actions for sample preparation are taken according to characteristics of samples.In this way,the accurate testing-results are ensured,and the average usage times of the combustion tube and reduction tube have been increased by 24% and 59%,respectively.Correspondingly,the efficiency of element analyzer is enhanced.
作者 吴冰 曹向前
出处 《苏州大学学报(自然科学版)》 CAS 2010年第3期87-90,共4页 Journal of Soochow University(Natural Science Edition)
关键词 元素分析 燃烧管 还原管 填充 制样 element analysis combustion tube reduction tube filling sample preparation
  • 相关文献

参考文献6

二级参考文献16

  • 1Zech W, Senesi N, Guggenberger Get al. Geoderma, 1997,79(1--4) :117-161
  • 2Solomon D, Fritzsche F, Lehmann J et al. Soil Sci Soc Am J,2002,66(3):969-978
  • 3Spaccini R,Mbagwu J S C,Igwe C A et al. Soil Tillage Re, 2004,75(2) : 161-172
  • 4余仲建,有机元素定量分析,1959年
  • 5(SH/T0656 1998)中华人民共和国石油化工行业标准
  • 6ASTMD5291-94美国材料与试验协会标准
  • 7Steyermark A L.Quantitative organic microanalysis[M].New York:Academic Press,1961:222.
  • 8Bishara S W,Attia M E,Hassan H N A.A new method for the microdetermination of carbon and hydrogen in fluoro and perfluoro compounds[J].Microchimica Acta,1974,62:819.
  • 9Alicino J F.The use of vermiculite in carbon and hydrogen determinations[J].Microchem J,1965,9:22.
  • 10Campbell A D,Macdonald A M G.Determination of carbon and hydrogen in organic fluorine compounds[J].Anal Chim Acta,1962,26:275.

共引文献51

同被引文献25

引证文献4

二级引证文献20

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部