数字化X射线成像系统MTF的刀口测量法 被引量:7

Edge Measurement for Modulation Transfer Function in Digital Radiography X-Ray Imaging System
下载PDF
导出
摘要 讨论了数字化X射线成像系统调制传递函数(MTF)的刀口测量法.首先通过数值模拟方法研究了采样率和噪声对准确测量MTF的影响,真实测量中采用多条过采样刀口响应函数(ERF)取平均以压制噪声获取光滑的ERF曲线;然后用指数函数对所获ERF进行分段拟合以确保刀口边缘强度值变化的单调性,将由其获得的MTF值与线对卡结果进行比较,证明了该方法用于成像系统MTF测量的有效性.同时,通过实验分析介绍了刀口法材料、材料厚度、光源曝光电压及曝光时间的选择,最终在本系统测量中选择的刀口材料为1 mm厚铅板,曝光管电压为33 kV,曝光时间为10 s. The effectiveness of edge measurement for the modulation transfer function (MTF)in the digital radiogra- phy X-ray imaging system has been discussed. Firstly, numerical simulation was performed to study the effect of sampling rate and noise on precise measurement of MTF. Then, a smoother curve of edge respond function (ERF) was obtained by averaging some over-sampled ERFs to suppress the noise. Finally, exponential curve-fitting was applied to different segments to keep the monotonicity of ERF. A great consistency was achieved between the MTF of the test card and that obtained with the proposed method, which verifies the effectiveness of the method for MTF measure- ment in the imaging system. The choice of edge material, the thickness of material, the exposure voltage, as well as the exposure time was analyzed through experiment and at last, a lead sheet with the thickness of 1 mm was utilized under the exposure voltage of 33 kV and the exposure time of 10 s in the proposed measurement.
出处 《天津大学学报》 EI CAS CSCD 北大核心 2010年第7期611-618,共8页 Journal of Tianjin University(Science and Technology)
基金 国家自然科学基金资助项目(30870657 30970775) 国家高技术研究发展计划(863计划)资助项目(2009AA02Z413) 国家重点基础研究发展计划资助项目(2006CB705700) 天津市科学技术委员会基金资助项目(09JCZDJC18200)
关键词 调制传递函数 有效性 刀口法 空间分辨率 modulation transfer function (MTF) effectiveness edge spatial resolution
  • 相关文献

参考文献14

  • 1Fujita H,Tsai D Y,Itoh T,et al. A simple method for determining the modulation transfer function in digital radiography[J].IEEE Transactions on Medical Imaging,1992,11 (1) 34-39.
  • 2康立丽,邓晓刚,陈阳,杨绍洲.影像系统MTF算法实现及分析[J].北京生物医学工程,2002,21(3):179-179. 被引量:13
  • 3梁丽红,路宏年.X射线数字成像系统传递函数模型[J].北京航空航天大学学报,2005,31(9):1018-1021. 被引量:4
  • 4叶昶,仲爱军,骆清铭.用于小动物研究的数字X线成像系统[J].中国医疗器械杂志,2004,28(3):179-181. 被引量:2
  • 5周俸才,管永红.闪光X射线照相系统调制传递函数测量[J].光子学报,2001,30(2):218-224. 被引量:8
  • 6Samei E, Flynn M, Reimann D. A method for measuring the presampled MTF of digital radiographic systems using an edge test device [J]. Medical Physics, 1998, 25 (1) : 102-113.
  • 7Illers H,Buhr E,Gunther-Kohfahl S,et al. Measurement of the modulation transfer function of digital X-ray detectors with an opaque edge-test device [J]. Radiation Protection Dosimetry,2005,114( 1/2/3) 214-219.
  • 8Zhu Yuemin, Kaftandjian Valerie ,Peix Gilles ,et al. Modulation transfer function evaluation of linear solid- state X-ray-sensitive detectors using edge techniques [J]. Appl Opt, 1995,34 (22) :4937-4943.
  • 9Tatsuya Yamazaki,Makoto Nokita,Shinsuke Hayashida, et al. A method to measure the presampling MTF using a novel edge test device and algorithm [C]//Medical Imaging 2004: Physics of Medical Imaging. San Diego,CA, USA : SP1E, 2004 : 696-704.
  • 10Cunningham I A, Reid B K. Signal and noise in modulation transfer function determinations using the slit, wire, and edge techniques [J].Medical Physics, 1992,19 (4) : 1037-1044.

二级参考文献17

  • 1[2]Judy PF. The line spread function and modulation transfer function of a computed tomographic scanner.Medical Physics,1976,3(4):233-236
  • 2[3]Shashadhar MM, Jeff DT, et al. Transfer function measurement and analysis for a magnetic resonance imager. Medical Physics,1991,18(6):1141-1144
  • 3[4]Michael CS, Dick JD. Frank P.Computing the modulation transfer function of a magnetic resonance imager. Medical Physics,1994,21(3):483-489
  • 4胡曙光,昌仁民,冯小刚,康立丽,林意群,张雪林.CT的MTF测量和评价[J].中国医学物理学杂志,1997,14(1):30-32. 被引量:7
  • 5R.Fahrig et al.X-ray imaging using amorphous selenium Detective quantum efficiency of photoconductive image receptors for digital mammography.Med.Phys.1995,22(4):153-160
  • 6R.M.Nishikawa et al.Scanned projection digital mammography.Med.Phys.1987,14(5):717-727
  • 7H.Liu et al.Design of a dual CCD configuration to improve the signal-to-noise ratio.Med.Phys.2000,27(10):2435-2437
  • 8Hiroshi Fujita et al A simple method for determining the modulation transfer function in digital radiography.IEEE Trans Med Imag.1992,11(1):34-39.
  • 9Samei et al A method for measuring the presampled MTF of digital radiographic systems using an edge test device.Med.Phys.1998,25(1):102-113.
  • 10Fabrice Ouandji et al.Characterization of a CCD-based digital X-ray imaging system for small-animal studies properties of spatial resolution.2002,41(13):2420-2427.

共引文献22

同被引文献45

  • 1梁丽红,路宏年.X射线数字成像系统传递函数模型[J].北京航空航天大学学报,2005,31(9):1018-1021. 被引量:4
  • 2李旭东,惠渭生,胡铁力,付建明.红外热成像系统调制传递函数(M TF)测试研究[J].应用光学,2006,27(4):323-326. 被引量:12
  • 3戴奇燕,夏德深,何红艳,满益云,宗云花.刀刃法在轨MTF测量性能分析[J].航天返回与遥感,2006,27(3):22-27. 被引量:17
  • 4黄巧林,姜伟.航天光学遥感器MTF测试技术研究[J].航天返回与遥感,2006,27(4):33-37. 被引量:8
  • 5喻春雨,常本康,魏殿修.新型X光成像系统及其性能分析[J].仪器仪表学报,2007,28(1):150-153. 被引量:4
  • 6Illers H,Buhr E,Gunther-Kohfahl S,et al.Measurement of the modula-tion transfer function of digital X-ray detectors with an opaque edge-test device[J].Radiation Protection Dosimetry,2005,114(1/2/3):214-219.
  • 7Wells J R, Dobbins J TIII. Estimation of the two-dimen- sional presampled modulation transfer function of digital ra- diography devices using one-dimensional test objects [ J ]. Med Phys, 2012, 39(10): 6148.
  • 8Illers H, Buhr E, Gunther-Kohfahl S, et al. Measurement of the modulation transfer function of digital X-ray detectors with an opaque edge-test device [ J ]. Radiation Protection Dosimetry, 2005, 114 (1/2/3) : 214-219.
  • 9Nandini N, Murthy K S, Kumar G H. Estimation of skew angle in binary document images using Hough transform [ C]//Proceedings of World Academy of Science. 2008, 42: 44-49.
  • 10Greer P B, van Doom T. Evaluation of an algorithm for the assessment of the MTF using an edge method. Med Phys, 2000, 27 (9) : 2048-2059.

引证文献7

二级引证文献44

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部