摘要 文中提出一个FAIMS仪用高场非对称方波RF电源系统设计方案,系统由信号发生器、高速开关、高压直流电源、高通滤波、补偿电压自动扫描及采集、和计算机控制六部分组成。以TMS320F2812为信号发生及控制核心,驱动高速开关通过逆变产生方波RF信号同时控制DAC转换实现补偿电压自动扫描。该电源系统可以输出频率≤2MHz,电压幅值≤2kV,功率≤300W,占空比20%~50%,波形上升时间约15ns的非对称方波波形。 A high-field high-frequency rectangular waveform RF power supply for FAIMS (field asymmetric waveform ion mobility spectrometry)system is developed in this paper. The system consists of six parts, which are signal generator, high-speed switch, high-voltage DC power supply, high-pass filter, the compensation voltage auto-sweeping and collection and computer control circuit. TMS320F2812 generaled and controlled the signals, which drivered the highspeed switch to general the rectangular RF waveform by inverter. The power system could general rectangular asymmetric waveform, limited to frequencies of ≤2MHz, voltage pulse amplitudes of ≤2kV, power of ≤300W and the duty cycle of 20%-50% with waveform steep rise times of the order of -15ns. The toluene sample was adopted to characterize the FAIMS system. When the power system was applied to the FAIMS system, the spectra of toluene were obtained, it shows that the ions can be separated and detected by the FAIMS system.
出处 《自动化与仪表》 北大核心 2010年第9期13-16,共4页 Automation & Instrumentation
基金 国家自然科学基金(60706030) 国家自然科学基金仪器专项(20827007) 国家863计划项目(2007AA04Z337) 国家973计划项目(2007CB310504)资助
关键词 TMS320F2812 高场非对称波形离子迁移谱 信号发生器 PWM 高场非对称方波 射频电压 TMS320F2812 high-field asymmetric waveformion mobility spectrometry (FAIMS) signal generator PWM high-filed asymmetric rectangular waveform RF voltage
  • 相关文献

参考文献3

二级参考文献23

  • 1杜华.任意波形发生器及其应用[J].国外电子测量技术,2005,24(1):38-40. 被引量:11
  • 2邢喆,朱绣鑫.多功能信号发生器[J].电子测量技术,2004,27(6):18-19. 被引量:4
  • 3Earl Wadsworth McDaniel, Edward A Mason. The mobility and diffusion of ions in gases[M]. New York, USA: Willey, 1973.
  • 4Shvartsburg Alexandre A, Keqi Tang, Smith Richard D. Optimization of the design and operation of FAIMS analyzers[J]. J Am Soc Mass Spectrom, 2005, 16(1): 2-12.
  • 5Eiceman G A, Nazarov E G, Miller R A. A micro-machined ion mobility spectrometer-mass spectrometer[J]. IJIMS, 2000, 3 (1):15-27.
  • 6Raanan A Miller, Gary A E;iceman, Erkinjon G Nazarov. A micromaehined field asymmetric ion mobility spectrometer (FAIMS)[J]. IJIMS, 2002, 5(2): 118-134.
  • 7Krebs M D, Zapata A M, Nazarov E G, et al. Detection of biological and chemical agents using differential mobility spectrometry (DMS) technology[J]. Sensors Journal, 2005, 5 (4): 696- 703.
  • 8Borsdorf H, Nazarov E G, Miller R A. Time of-flight ion mob ility spectrometry and differential mobility spectrometry: A comparative study of their efficiency in the analysis of halogenated compounds[J]. Talanta, 2007, 71(4): 1804-1812.
  • 9Guevremont Roger, Purves Randy. Comparison of experimental and calculated peak shapes for three cylindrical geometry FAIMS prototypes of differing electrode diameters[J]. J Am Soc Mass Spectrom, 2005, 16(3): 349 362.
  • 10Miller R A, Eiceman G A, Nazarov E G, et al. A novel micromachined high field asymmetric waveform-ion mobility spectrometer[J]. Sensors and Actuators B, 2000, 67(3): 300-306.