期刊文献+

便携式HGB测量系统设计

Design of Portable HGB Measurement System
下载PDF
导出
摘要 针对目前单波长法HGB(血红蛋白)测量仪需要测量空白溶液进行对比校正的缺陷,提出了一种便携式双波长HGB测量系统。系统主要包括恒定光源、比色池、光接收器、ARM控制与数据处理单元、LCD显示等模块。经过实际测试表明,该系统性能稳定,精度高,比单波长法更加高效,可以较好的满足临床医学的需要,具有一定的实用性和推广价值。 This paper proposes a portable dual-wavelength HGB measurement system to avoid the defect that the single-wavelength HGB (Hemoglobin) admeasuring apparatus needs to measure the blank solution for comparison and correction,The system consists of constant light sources,residual,core controller based on ARM,data processing unit,LCD part and so on. Compared with the methods of single-wavelength measurement,the system is proved to be high-efficiency and can work at a high speed,and well meet the need of clinical medicine,The technology and methods adopted in the system are practical and can be extensively used.
出处 《中国医疗器械杂志》 CAS 2010年第5期343-346,共4页 Chinese Journal of Medical Instrumentation
基金 浙江省科技厅科技计划项目(2009C31003) 宁波市服务型重点专业建设项目(010-A00083094702)
关键词 双波长法 嵌入式 血红蛋白 S3C44B0X Measurement of Dual-wavelength Embedded Technology Hemoglobin S3C44B0X
  • 相关文献

参考文献4

  • 1Keiji Fujimoto,Principle of measurement in hematology analyzers manufactured by sysmex corporation[J].Sysmex Journal International,1999,9(1):43-60.
  • 2http://www.samsung.com/Products/Semiconductor/MobileSoC/ApplicationProcessor/ARM7Series/S3C44B0/S3C44B0.htm,SUMSANG Co.,Ltd.,2005.
  • 3孙肖子,杨颂华.数字电子技术基础[M].电子科技大学出版社,2000.
  • 4刘广.介绍一种人工校准血液分析仪的简单方法.陕西医学检验,1996,11(2):48-48.