低管电压对双源CT冠状动脉成像图像质量和辐射剂量的影响 被引量:26

The effect of low tube voltage on image quality and radiation dose with dual-source coronary CT angiography
下载PDF
导出
摘要 目的:分析低管电压双源CT冠状动脉成像(CTA)的图像质量,探讨低管电压扫描对减少冠状动脉CTA检查辐射剂量的价值。方法:150例患者分成两组行双源CT冠状动脉CTA检查,A组使用100kV管电压扫描,B组使用120kV管电压扫描。比较两组的辐射剂量,同时分析并比较两组冠状动脉的整体图像质量。结果:A、B两组的辐射剂量分别为(6.80±1.70)mSv和(13.01±2.61)mSv,A组辐射剂量明显小于B组,差异有统计学意义(t=17.168,P=0.000)。A、B两组冠状动脉图像质量评分分别为(2.87±0.25)分和(2.92±0.17)分,两组间差异无统计学意义(Z=-1.067,P=0.286)。结论:采用100kV管电压扫描行冠状动脉CTA检查能明显减小正常和略胖体型患者所接受的辐射剂量,同时图像质量不受影响。 Objective:To study the image quality of low tube voltage with dual-source coronary CT angiography (CTA) and to investigate the application of low tube voltage for reducing radiation dose.Methods:150 patients underwent dual-source coronary CTA were divided to group A and B with the tube voltage as 100kV and 120kV,respectively.The radiation dosage and image quality of coronary CTA between group A and group B were compared.Results:The effective radiation dose was (6.80±1.70)mSv for group A and (13.01±2.61)mSv for group B.The effective radiation dose for group A was obviously lesser than that for group B (t=17.168,P=0.000).The image quality score of coronary artery for group A and B was 2.87±0.25 and 2.92±0.17,respectively.There was no statistic difference between these two groups (Z=-1.067,P=0.286).Conclusion:Using the tube voltage as 100kV can obviously reduce the radiation dose for patients with normal or slight obese figure while with no influence on image quality.
机构地区 解放军第
出处 《放射学实践》 北大核心 2010年第9期999-1002,共4页 Radiologic Practice
关键词 冠状动脉 体层摄影术 X线计算机 管电压 辐射剂量 Coronary artery Tomography X-ray computed Tube voltage Radiation dose
  • 相关文献

参考文献5

二级参考文献31

共引文献296

同被引文献247

引证文献26

二级引证文献154

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部