摘要
目的对我区4所二级医院的4台血细胞分析仪(2台型号相同,其余2台型号不同)测定值的准确性进行评估,探讨不同仪器测定结果间差异是否有统计学意义。方法应用3份抗凝新鲜全血由4台仪器在1个小时内测定,计算其测定值CV和极差。结果不同仪器测定值最大变异系数比较:WBC(2.7%);RBC(1.0%);HGB(2.7%);PLT(3.8%);HCT(1.4%)。结论开展实验室质量管理工作,提高全血细胞计数检测质量水平,通过比对试验,控制好不同仪器检测结果之间的差异,使同一病人标本在不同医院的测定结果趋向一致,是今后工作中应重点解决的问题,为检验结果互认奠定良好的基础。
Objective 4 2 hospitals in our region 4 hematology analyzer (2 sets of the same model,and the remaining 2 sets of different models) to assess the accuracy of measured values,the results of determination of different instruments are statistically significant differences between.Methods 3 copies of fresh whole blood coagulation instrument from the 4 in 1 hour and measured to calculate the measured values of CV and range.Results different instruments maximum variation coefficient of determination values:WBC(2.7%);RBC(1.0%);HGB(2.7%);PLT(3.8%);HCT(1.4%).Conclusion laboratory quality management to improve the quality of complete blood cell count level,through the comparison test,control,detection of different instruments well the difference between the results,make the same patient specimen results in different hospitals in the determination of consistency,is the future work should Key issues for mutual recognition of test results lay a good foundation.
出处
《当代医学》
2010年第27期12-13,共2页
Contemporary Medicine
关键词
血细胞计数
自动分析
比较研究
Blood cell count
automatic analysis
Wuality control
Comparative study