期刊文献+

针对小动物的锥束CT重建的研究及验证 被引量:4

Investigation and Validation of the Reconstruction Algorithm for Small Animal Cone-beam CT Imaging
下载PDF
导出
摘要 基于经典3D Shepp-Logan模型,应用射线追踪和射线衰减定律得到锥束CT的模拟仿真数据.利用FDK三维重建算法实现锥束CT图像重建,采用仿真图像对重建算法进行了验证,通过搭建的锥束CT系统和自制仿体进行了层析成像和图像重建.结果表明:算法可较为准确地重建出原物体的空间分布和密度关系,针对仿真数据的中心断层重建图像的空间分辨率可达到8lp/mm. Simulation for cone beam CT system and the 3D Shepp-Logan Model is carried out.FDK reconstruction algorithm is developed to reconstruct the images and then evaluated with the 3D Shepp-Logan Mode.Experiments are carried out on a constructed cone beam CT system and a homemade phantom.Simulation and experimental reconstruction results indicate that reconstruction algorithm can faithfully provide both the position and relative contrast of the inner objects.The spatial resolution of the central image from simulation data can reach 8 linepairs/mm.
出处 《光子学报》 EI CAS CSCD 北大核心 2010年第9期1588-1593,共6页 Acta Photonica Sinica
基金 国家自然科学基金(30870657) 天津市重点基金(09JCZDC18200) 国家重点基础研究发展规划(2006CB705700)资助
关键词 锥束CT 重建算法 FDK Cone-beam CT Reconstruction algorithm FDK
  • 相关文献

参考文献10

  • 1HSIEH J.Computed tomography:principles,design,artifacts,and recent advances[M].Bellingham:SPIE,2002.
  • 2ZHU Shou-ping,TIAN Jie,YAN Guo-rui,et al.Cone beam micro-CT system for small animal imaging and performance evaluation[J].Int J Biomed Imaging,2009,2009:960573-960579.
  • 3ZHU Shou-ping,TIAN Jie,YAN Guo-rui.An experimental cone-beam micro-CT system for small animal imaging[C].SPIE,2009,7258:72582S.
  • 4孙丰荣,刘泽,李艳玲,曲怀敬,张梅,张运.基于模型的CT三维医学图像重建仿真[J].系统仿真学报,2006,18(3):781-784. 被引量:14
  • 5SMITH B D.Image reconstruction from conebeam projections:necessary and sufficient conditions and reconstruction methods[J].IEEE Trans Med Imaging,1985,MI-4(1):4-25.
  • 6KAK A C,SLANEY M.Principles of computerized tomographic imaging[M].New York:IEEE,1988.
  • 7叶志强,吴建明.基于CT切片特征匹配的三维体重建(英文)[J].光子学报,2003,32(1):22-25. 被引量:3
  • 8CHEN Zi-kuan,CALHOUN V D,CHANG Sheng-jiang.Compensating the intensity fall-off effect in cone-beam tomography by an empirical weight formula[J].Appl Opt,2008,47(32):6033-6039.
  • 9GRASS M,KOHLER T,PROKSA R.3D cone-beam CT reconstruction for circular trajectories[J].Phys Med Biol,2000,45(2):329-347.
  • 10TURBELL H.Cone-beam reconstruction using filtered backprojection[D].Link(o)poing:Link(o)poing University,2001.

二级参考文献15

  • 1L A Shepp, B F Logan. Reconstruction interior head tissue from x ray transmissions [J].IEEE Trans. Nuclear Sci. (S0018-9499), 1974,NS-21:228-236.
  • 2C R Crawford, K F King. Computered Tomography Scanning with Simultaneous Patient Translation [J]. Med. Phys. (S0094-2405), 1990,17(6): 967-982.
  • 3J P Heiken, J A Brink, M W Vannier. Spiral (Helical) CT [J].Radiology (S0033-8419), 1993, 189(3): 647-656.
  • 4Jiang Hsieh. A General Approach to the Reconstruction of x-ray Helical Computed Tomography [J]. Med. Phys. (S0094-2405), 1996,23(2): 221-229.
  • 5Harish P. Hiriyannaiah. X-ray Computed Tomography for Medical Imaging[J]. 1EEE Signal Processing Magazine (S1053-5888), 1997,14(1): 42-59.
  • 6Ying Liua, Hong Liu, Ying Wang, G-e Wang. Half-scan cone-beam CT fluoroscopy with multiple x-ray sources [J]. Med. Phys. (S0094-2405),2002, 28(7): 1466-1471.
  • 7Anil K Jain. Fundamentals of digital image processing [M]. NJ:Prentice-Hall, 1989.
  • 8Pardo X M, Carreira M J, MosqueraA, et al. A snake for CT image segmentation integrating region and edge information. Imageand Vision Computing, 2001, 19: 461~475
  • 9Martin O, Emmanuel B, Emmanuelle C. Two-photo microscopy in brain tissue:parameters influencing the imaging depth. Journal of Neuroscience Methods Ⅲ, 2001, 3: 29~37
  • 10Gambarini G, Gomarasca G. Three-dimensional determination of absorbed dose byspectrophotometric analysis of ferrous-sulphate agarose gel. Nuclear and Methods inPhysics Research (A), 1999, 422: 643~648

共引文献15

同被引文献39

  • 1祝海江,吴福朝.基于一组对应消失线的度量重建[J].软件学报,2004,15(5):666-675. 被引量:3
  • 2GULLBERG G T,TSUI B M W, CRAWFORD C R,et al. Estimation of geometrical parameters and collimator evalu- ation for cone-beam tomography [ J ]. Medical. Physics, 1990,17(2) :264-272.
  • 3LI J Y, JASZCZAK R J, WANG H L, et al. Determina- tion of both mechanical and electronic shifts in cone beam SPECT [ J ]. Physics in Medicine and Biology, 1993,38 (6):743-754.
  • 4RIZO P, GRANGEAT P, GUILLEMAUD R. Geometric calibration method for multiple-head cone-beam SPECT system[ J]. IEEE Trans. on Nuclear Science, 1994,41 (6) :2748-2757.
  • 5NOO F, CLACKDOYLE R, MENNESSIER C, et at. Analytic method based on identification of ellipse parameters for scanner calibration in cone-beam tomography[J].Physics in Medicine and Biology ,2000,45 ( 11 ) :3489-3508.
  • 6ROBERT N,WATI" N K,WANG X Y ,et al. The geometric cal- ibration of cone-beam systems with arbitrary geometry [ J ]. Physics in Medicine and Biology,2009,54(24) :7239-7261.
  • 7SUN Y,HOU Y,ZHAO F Y,et al. A calibration method mis- aligned scanner geometry in cone-beam computed tomo- graphy[J]. NDT and International,2006,39(6):199-513.
  • 8SUN Y,HOUY,HU J SH. Reduction of artifacts induced by misaligned geometry in cone-beam CT [ J ]. IEEE Trans. onBiomedical Engineering ,2007,54 ( 8 ) : 1461-1471.
  • 9闫镔,杜剑平,李建新,等.VCT系统参数标定装置及方法:中国,ZL2009101623297[P].2009-12-30.
  • 10KINGSTON A M, SAKELLARIOU A,SHEPPARD A P,et al. An auto-focus method for generation sharp 3D tomo- graphic images [ J ]. Proe. of SPIE, 1804:78040J.