LC-MS/MS法和mFPIA法测定全血环孢素A浓度的相关性比较

Correlations of LC-MS/MS and mFPIA for the Determination of Whole Blood Cyclosporin A Concentration
原文传递
导出
摘要 目的:对高效液相色谱电喷雾串联质谱(LC-MS/MS)法和单克隆荧光偏振免疫(mFPIA)法测定肾移植术后稳定期患者不同时间点的全血环孢素A(CsA)浓度进行相关性比较。方法:分别采用LC-MS/MS法和mFPIA法测定患者全血CsA浓度,对测定结果进行双侧配对t检验比较差异性,用Passing-Bablok回归分析法比较不同采血时间点2种方法测定结果的相关性。结果:2种方法测定值之间具有显著性差异;服药后0、3.0、4.0h时2法测定值的相关性最好,0.5、0.75、12.0h相关性最差。结论:对CsA血药浓度监测结果的分析需考虑监测方法以及不同采血时间点的影响。 OBJECTIVE:To compare the correlations of LC-MS/MS and mFPIA for the determination of the whole blood Cyclosporin A(CsA)concentration in patients after renal transplantation in stable phase at different time.METHODS:The concentrations of CsA in patients were assayed by LC-MS/MS and mFPIA respectively.The difference of results was analyzed by the two-sided paired t-test and the correlation between the results of 2 kinds of methods at different time were detected by Passing-Bablok regression analysis.RESULTS:There was significant difference between results of 2 kinds of methods;obvious correlation was found between results of 2 kinds of methods at 0 h,3.0 h,4.0 h after administration and no obvious correlation was found at 0.5 h,0.75 h,12.0 h.CONCLUSION:Effect of monitoring method and different sampling time should be taken into account in monitoring analysis of blood concentration of CsA.
出处 《中国药房》 CAS CSCD 北大核心 2010年第38期3593-3595,共3页 China Pharmacy
基金 山东省医药卫生科技发展计划项目(2007HZ011) 2009年度山东省药学会临床药学奥赛康中青年科研资助项目
关键词 高效液相色谱电喷雾串联质谱法 单克隆荧光偏振免疫法 相关性 环孢素A 全血浓度 LC-MS/MS mFPIA Correlation Cyclosporin A Whole blood concentration
  • 相关文献

参考文献9

二级参考文献24

  • 1林发全,李明芬.不同血凝分析仪的检测结果可比性探讨[J].第三军医大学学报,2005,27(17):1803-1804. 被引量:11
  • 2毛桂福.高效液相色谱法测定人全血中环孢素A浓度[J].中国医院药学杂志,2005,25(5):420-422. 被引量:11
  • 3方芸,王欣,裴云萍.环孢素A监测指标及治疗浓度的研究进展[J].中国药房,2006,17(4):307-309. 被引量:10
  • 4杨晓英.电泳法与化学法测定血清白蛋白、白蛋白/球蛋白比值结果探讨[J].检验医学,2006,21(4):350-351. 被引量:4
  • 5JOEL G. H, LEE E. Goodman & Gilman 'S Pharmacological Basis of Therapeutics(古德曼吉尔曼药动学的药理学基础)[M].10th ed. Bejing: People's Medicinal Press. 2002:1339.
  • 6NAPOLI K L. 12-hour area under the curve cyclosporine concentrations determined by a validated liquid chromatographyspectrometry procedure compared with fluorescence polarization immunoassay reveals sirolimus effect in cyclosporine pharmacokinetics[J]. Ther Drug Monit, 2006, 28(6):726-735.
  • 7CHENJS DUQY.Cyclosporin A in whole blood drug concentration monitoring method evaluation and its clinical significance.中国医院药学杂志,2004,46(8):486-487.
  • 8SALM P, TAYLOR P J, LYNCH S V, et al. A rapid HPLC-mass spectrometry cyclosporin method suitable for current monitoring practices[J]. Clin Biochem, 2005, 38(7):667-673.
  • 9TAYLOR P J, JONES C E, MARTIN P T, et al. Microscale high performance liquid chromatography- electrospray mass spectrometry assay for cyclosporin A in blood[J]. J Chromatogr B Biomed Sci Appl, 1998,705(2):289-294.
  • 10MENGXM YINJ ZHANGH etal.Cyclosporine in the neural immune disease in the clinical application.中国神经免疫学和神经病学杂志,2006,16(1):22-24.

共引文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部