摘要 数字化乳腺X线摄影技术已成为临床检测早期无症状隐匿性乳腺癌的重要手段,在此背景下,评价数字乳腺机性能的量子探测效率(DQE)检测方法,是国内外临床医学界和医学影像检测部门所关注的重要课题。针对相对成熟的IEC62220-1标准不适用于辐射剂量小、辐射野小的数字乳腺机DQE检测的问题,本文较深入地研究了基于IEC62220-1-2标准的数字乳腺机DQE检测方法和具体数字图像处理算法的实现。并将该DQE检测手段实施于辽宁省医疗器械检验所受检产品某公司生产的乳腺机,得到了相应的DQE测试结果,最后,本文讨论了数字化乳腺成像系统量子探测效率的重要影响因素。 Quantum detective efficiency (DQE) measurement methodology based on the international standard IEC 62220-1 is being regarded as an objective performance evaluation metric for the digital X-ray graphic system. However, the fast becoming popular mammographic systems are not applied to the IEC 62220- 1 standard because the factors affecting measurement are more complicated. In this paper, the DQE measurement methodology for digital mammographic system and the corresponding standard IEC 62220-1-2 was investigated. Further, this measurement method was used for testing the DQE of the digital mammographic system. The DQE value was obtained and the factors affecting measurement were discussed in detail.
出处 《北京生物医学工程》 2010年第5期441-448,共8页 Beijing Biomedical Engineering
基金 国家自然科学基金(60771067)资助
关键词 数字乳腺机 量子探测效率 边缘检测 调制传递函数 噪声功率谱 digital mammographic system detective quantum efficiency edge detection modulation transfer function noise power spectrum,
  • 相关文献

参考文献7

  • 1IEC 62220-1.Medical electrical equipment – Characteristics of digital X-ray imaging devices – Part 1:Determination of the detective quantum efficiency.Geneva,Switzerland:International Electrotechnical Commission (IEC),2003.
  • 2IEC/CEI 62220-1-2.Part 1-2:Determination of the detective quantum efficiency –detectors used in mammography.First edition.2007.
  • 3Park HS,Kim HJ,Cho HM,et al.Measurements and Evaluation of the Image Noise Power Spectrum for Computed Radiography.IEEE,2008,M06-459.
  • 4Buhr E,Gunther-Kohfahl S,Neitzel U.Simple method for modulation transfer function determination of digital imaging detectors from edge images.Proceeding of SPIE,2003,5030:877-884.
  • 5Zhang D,Rong J,Chu R,et al.DQE measurements in magnification X-ray Imaging.Journal of X-Ray Science and Technology,2006,14:141-150.
  • 6Takamura M,Higaki A.Analysis of factors that affect DQE in digital mammography.Conference on Physic of Medical Imaging.Proceedings of SPIE,2004,5745:1217-1228.
  • 7白玫,彭明辰.X线CT成像质量评估中的调制传递函数的测试与分析[J].医疗设备信息,2006,21(5):11-13. 被引量:15

二级参考文献7

  • 1Goodenough D,Atkins F,Dyer S.Automated techniques for quality assurance of radiological image modalities.SPIE.Image Capture,Formatting,and Display,1991,1444:87-99.
  • 2Dainty JC,Shaw R.Image Science[M].New York:Academic Press,1974.
  • 3Goodenough D,Atkins F.Theoretical and practical aspects of automated quality assurance approaches,particularly for CT[C].Proceedings of the ICR 89,editors Silve,et.Al,1990.
  • 4Fujita H,Tsai D-Y,Itoh T,et al.A simple method for determining the modulation transfer function in digital radiography[J].IEEE Transactions on Medical Imaging,1992,11:34-39.
  • 5Cunningham IA,Reid BK.Signal and noise in modulation transfer function determinations using the slit,wire,and edge techniques[J].Medical Physics,1992,19:1037-1044.
  • 6Samei E,Flynn MJ,Reimann DA.A method for measurement the presampled MTF of digital radiographic systems using an edge test device[J].Medical Physics,1998,25:102-113.
  • 7陈欠根,陈大舵.CT图像中轮廓特征的提取[J].医疗设备信息,2003,18(1):10-12. 被引量:3