期刊文献+

X射线衍射仪故障与维修实例 被引量:2

Troubleshooting and Servicing Example of X-ray Diffraction
下载PDF
导出
摘要 本文研究了X射线衍射仪的样品台的驱动原理与机械结构设计。通过对设备构造的分析,成功的更换了X射线衍射仪的主机电源,更换了L6202电机驱动模块,修复了控制样品台Y轴电机的驱动电路。最后排除了导致设备短路的故障点。此次维修为大型仪器设备的维护积累了宝贵的经验。 In this paper, the driving principle and the mechanical structure of the specimen holder of the X-ray diffraction is investigated. Through the analysis of the equipment construction, the power supply of the X-ray diffraction was successfully replaced. And then we replaced the L6202 motor driving and repaired the Y-axis motor driving circuit of the specimen holder. At last, short circuit points leading to equipment failure were eliminated. This servicing accumulates valuable experience for the maintenance of laree eouinment.
作者 刘伟 周轶然
出处 《现代科学仪器》 2010年第5期143-144,148,共3页 Modern Scientific Instruments
关键词 X射线衍射仪 样品台 L6202电机驱动模块 X-ray diffraction Specimen holder L6202 motor driving module