非侵入性聚焦超声融脂的临床疗效观察 被引量:8

Clinical Observation on Body Contouring by Non-Invasive Focused Ultrasound
下载PDF
导出
摘要 目的:为观察非侵入性聚焦超声融脂机的临床疗效。方法:应用以色列Ultrashape公司生产的聚焦超声ContourI塑形治疗仪,以功率密度17.5W/cm2,声波频率(200±30)KHz,脉冲周期400ms,治疗用声波输出持续时间1s,对38例患者的腹部皮下脂肪进行减脂塑形治疗。患者间隔两周治疗一次,最末次治疗后两周随访。结果:38例患者中,28例治疗了3次,6例治疗了4次,4例治疗了6次。治疗过程中,治疗部位未见皮肤红肿、水疱、溃疡。38例患者治疗前后体重平均下降(0.45±1.24)Kg,3次治疗腹围减少(2.20±1.94)cm,4次治疗腹围减少(2.92±2.31)cm,6次治疗腹围减少(4.75±2.52)cm。结论:应用聚焦超声融脂机进行减脂塑形治疗,临床证明有效,为临床提供了一种新的无创、非侵入性治疗方法。 Objective To observe the clinical effect of body contouring treated by non-invasive focused ultrasound. Methods The unwanted fat on abdomen of 38 patients were treated by an non-invasive focused ultrasound device [ UltraShape Contour I,Israel],These cases were cured by this system with power density 17.5W/cm2, acoustic frequency (200±30) KHz,pulse period 400ms,the duration of treatment with the acoustic output ls. The treatment in patient was with an interval of two weeks. The follow-ups were made after two weeks of last time, Results Among 38 patients,28 cases were treated for 3 times, 6 cases for 4 times,and 4 cases for 6 times. No skin irritation,blisters,and ulcers occurred in the treatment. The average body weight of 38 patients were decreased in (0.45±1.24) kg, The abdominal circumference of patients treated for 3 times were decreased in (2.20±1.94) cm,(2.92±2.31) cm for 4 times,and in (4.75± 2.52) cm for 6 times. Conclusion This clinical observation shows the efficacy of focused ultrasound machine for body contouring, and the therapeutic effect which can be improved by repeated treatment.
出处 《中国美容医学》 CAS 2010年第10期1438-1440,共3页 Chinese Journal of Aesthetic Medicine
关键词 塑形 非侵入性 聚焦超声 融脂 body contouring non-invasion focused ultrasound lipolysis
  • 相关文献

参考文献6

  • 1潘勇,韩岩,杨力,潘宝华,郭树忠.吸脂并发症及其防治[J].中国美容医学,2006,15(10):1201-1203. 被引量:3
  • 2胡志奇,高建华,齐向东.肿胀法行体外、内超声及单纯负压吸脂减肥术的比较[J].中华整形外科杂志,2002,18(4):221-223. 被引量:6
  • 3Teitelbaum SA,Bums JL,Kubota J,et al.Noninvasive body contouring by focused ultrasound:safety and efficacy of the contouring Ⅰ device in a Multicenter,controlled,clinical study[J].Plast Reconstr Surg,2007,120:779-789.
  • 4Commons GW,Halperin B,Cheng CC.Large volumes liposuction:Review of 631 consecutive cases over 12 years[J].Plast Reconstr Surg,2001,108:1753-1763.
  • 5Matarasso A.Swift RW,Rankin M.Abdominoplasty and abdominal contour surgery:A national plastic surgery survey[J].Plast Reconstr Surg,2006,117:1797-1808.
  • 6Moreno-Moraga J.Valero-Alté s T,Marti nez Riquelme A,et al.Body contouring by non-invasive transdermal focused ultrasound[J].Lasers in Surg and Med,2007,39:315-323.

二级参考文献6

  • 1Kenkel JM.The tissue effects of ultrasound-assisted lipoplasty[].Plastic and Reconstructive Surgery.1998
  • 2Thornton LK.Equipment and instrumentation for ultrasound-assisted lipoplasty[].Clinics in Plastic Surgery.1999
  • 3Igra H.Tumescent liposuction versus internal ultrasonic-assisted tumescent liposuction aside-to-side comparison[].Dermatologic Surgery.1997
  • 4Mustoe TA.Staff education and training for ultrasound-assisted lipoplasty[].Clinics in Plastic Surgery.1999
  • 5Klein JA.The tumescent for liposuction[].American Journal of Cosmetic Surgery.1997
  • 6Zoclchi M.Ultrasonic liposulpture[].Aesthetic Plastic Surgery.1992

共引文献7

同被引文献70

引证文献8

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部