摘要
目的观察和分析植入型心律转复除颤器(ICD)误放电的发生情况。方法对植入ICD患者进行随访,应用体外程控仪调出ICD存储的资料进行分析。结果31例植入ICD患者随访平均10个月中,2例患者因窦性心动过速发生ICD误放电8次。1例发生在运动过程中,1例发生于静脉滴注多巴酚丁胺时,经程控ICD的室性心动过速识别参数后,未再发生误放电。结论植入ICD的患者可能会发生误放电,定期的随访及调整合适的ICD参数,可避免和减少ICD误放电的发生。
Objective To observe and analyse the causes of inadequate shocks of ICD. Method Followup the patients with ICD implanted and analyse stored data of ICD. Results During the followup of a mean time of 10 months,the inadequate shocks were found in 2 patients out of 31 patients.Both patients had shocks induced by sinus tachycardia.In one patient inappropriate shocks happened during exercises and the other during intravenous infusion of dobutamine as a sort of treatment of congestive heart failure.There is no inappropriate shocks happened again after reprogramming detection parameters for VT of ICD. Conclusionlnadequate shocks could happen in patients with ICD implanted.Reprogramming appropriate detection criteria of ventricular tachycardia will avoid or reduce the occurrence of inadequate shocks.
出处
《中华心律失常学杂志》
1999年第2期103-105,共3页
Chinese Journal of Cardiac Arrhythmias
关键词
植入型
心律转复除颤器
窦性心动过速
误放电
Implantable cardioverter defibrillatorsSinus tachycardiaInadequate shock