摘要
目的:观察京尼平生物型人工食管的生物相容性。方法:应用急性毒性试验、亚急性毒性试验、溶血试验、细胞毒性试验、肌肉植入试验对材料的生物相容性进行评价。结果:材料无急性、亚急性毒性反应,溶血率为0.559%,低于5%,说明其有良好的血液相容性。细胞生长无明显抑制现象,材料细胞毒性评分为0级和1级。材料经肌肉植入1周后,形成纤维囊包裹,有少量淋巴细胞浸润;植入后4周材料周围出现较多淋巴细胞和嗜酸性粒细胞;植入后12周,材料周围淋巴细胞和嗜酸性粒细胞大量减少,组织炎症反应明显减轻。结论:京尼平生物型人工食管具有良好的生物相容性,符合临床应用前标准。
Objective: To investigate the biocompatibility of the genipin biotype artificial esophagus. Methods: The biocompatibility of the genipin biotype artificial esophagus was primarily evaluated using the acute and subacute toxicity tests, hemolysis test and cytotoxicity test and muscle implantation test. Results: No acute or subacute toxicity was revealed with the genipin biotype artificial esophagus; The hemolysis rate was 0. 559% (less than 5% ) ; The cytotoxicity scores were 0 and 1 grade. A mild inflammatory reaction with a small amount of lymphocytic infiltration was observed in rabbit' s muscle one week after implantation ; A large number of lymphocytes and eosinophils around the material were observed four weeks after implantation, but the inflammatory reaction reduced significantly 12 weeks after implantation. Conclusion: Our findings show that the genipin biotype artificial esophagus would be a promising biomaterial with good biocompatibility for clinical application.
出处
《肿瘤预防与治疗》
2010年第2期107-112,共6页
Journal of Cancer Control And Treatment
基金
四川省科技攻关资助项目(05SG022-022)
关键词
京尼平
人工食管
生物相容性
Genipin
Artificial Esophagus
Biocompatibility