期刊文献+

时间分辨免疫荧光与ELISA法检测乙型肝炎病毒血清标志物结果的比较与分析 被引量:2

An analysis and comparison on the detection of Hepatitis B virus serum markers between time-resolved immunofluorescence and ELISA
下载PDF
导出
摘要 目的:探讨时间分辨免疫荧光技术(TRFIA)定量检测乙型肝炎病毒标志物(HBV-M)在临床中的应用情况,并进行TRFIA的方法学研究.方法:实验用TRFIA法与ELISA法分别对200例血清标本进行HBsAg检测,对另200例血清样本进行"乙肝两对半"检测,同时检测20份HBsAg低值(弱阳性)标准品血清,并对TRFIA法与ELISA法进行结果比较与分析.结果:TRFIA法HBsAg灵敏度(0.2ng/ml)高于ELISA法,TRFIA法HBsAg线性检测范围在0.2-200ng/ml,且TRFIA法的特异性优于ELISA法,ELISA法在检测0.5-1ng/ml时易出现假阴性;且ELISA法易出现HBsAg、HBeAg或HBeAb假阴性,HBcAb假阳性,导致其他模式增多.结论:TRFIA法具有测量范围宽,灵敏度高、特异性强等优点,在临床检测HBV中具有广泛的应用前景. Objective: To investigate the clinical application of time-resolved immunofluorescence (TRF1A) used in HBV-M ration test and do some research on TRFIA method. Methods: To do HBsAg test 200 cases of serum samples HBsAg by TRFIA and ELISA and another 200 tested by HBV-M separately, at the same time 20 cases of standard blood in which HBsAg was tested and analyzes by ELISA and TRFIA. Results: TRFIA has higher sensitivity (0.2ng/ml) than ELISA. The specificity of TRFIA is superior to ELISA within the scope of linear test (0.2-200ng/ml), ELISA shows fake negative in testing in HBsAg, HBeAg or HBeAb, and fakes positive in HBcAb, and make other patterns increasing. Conclusion :TRFIA has higher specificity and sensitivity than ELISA.
出处 《中医临床研究》 2010年第22期55-56,共2页 Clinical Journal Of Chinese Medicine
  • 相关文献

参考文献4

二级参考文献5

共引文献74

同被引文献26

引证文献2

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部