摘要
In clinical practice,examination of the hemorrhagic spot (HS) remains difficult.In this paper,we describe a remote controlled capsule (RCC) micro-system with an automated,color-based sensor to identify and localize the HS of the gastrointestinal (GI) tract.In vitro testing of the detecting sensor demonstrated that it was capable of discriminating mimetic intestinal fluid (MIF) with and without the hemoglobin (Hb) when the concentration of Hb in MIF was above 0.05 g/ml.Therefore,this RCC system is able to detect the relatively accurate location of the HS in the GI tract.
In clinical practice, examination of the hemorrhagic spot (HS) remains difficult. In this paper, we describe a remote controlled capsule (RCC) micro-system with an automated, color-based sensor to identify and localize the HS of the gastrointestinal (GI) tract. In vitro testing of the detecting sensor demonstrated that it was capable of discriminating mimetic intestinal fluid (MIF) with and without the hemoglobin (Hb) when the concentration of Hb in MIF was above 0.05 g/ml. Therefore, this RCC system is able to detect the relatively accurate location of the HS in the GI tract.
基金
Project supported by the National Natural Science Foundation of China (Nos. 30700160 and 30970883)
the Postdoctoral Science Foundation of China (Nos. 20070420718 and 200801225)
Chongqing University Postgraduates’ Science and Innovation Fund (No. 2008 01A1B0250284)
the Specialized Research Fund for the Doctoral Program of Higher Education (No. 20070611045), China