高清晰CT胸部低剂量筛查的临床应用 被引量:11

Clinical application of the low-dose chest screening with high definition CT
下载PDF
导出
摘要 目的观察应用高清晰CT(HDCT)制订肺部低剂量筛查的扫描方案及可行性。方法将150名胸部体检者随机分为三组(试验组1、试验组2及对照组),应用HDCT,采用自动毫安调节技术行胸部CT扫描。试验组1、试验组2及对照组预设噪声指数(NI)分别为22HU、30HU及15HU,设置ASIR值分别为50%、50%和30%。记录三组的CT剂量指数(CTDIvol)、剂量长度乘积(DLP)并估算有效剂量(ED)。采用5分制对图像质量进行评估,并用Kappa检验评价观察者间的一致性。结果试验组1的CTDIvol为(3.00±1.41)mGy,试验组2为(1.48±0.53)mGy,对照组为(5.51±2.21)mGy。试验组1、试验组2的CTDIvol较对照组分别减少了45.55%和73.14%,DLP和ED分别减少了48.58%和73.72%。试验组2的ED降低至(0.71±0.28)mSv。三组图像的的质量均满足临床诊断要求,观察者间的评估具有一致性。结论采用HDCT,可以在获得稳定图像质量的前提下实现更低剂量的胸部CT筛查。 Objective To evaluate the program and the feasibility of low dose chest CT screening with high definition CT (HDCT). Methods Totally 150 subjects underwent chest screening using GE Discovery CT750 HD with automatic tube current (mA) modulation. All the subjects were devided randomly into three groups,including study group 1,2 and the control group. The noise index (NI) for the study group 1,2 and the control group was set to 22 HU,30 HU and 15 HU,respectively. The adaptive statistical interative reconstruction (ASIR) for each group was set to 50%,50% and 30%,respectively. CT dose index volumes (CTDIvol) and dose length product (DLP) were recorded,while effective dose (ED) was estimated. Image quality was assessed by two radiologists on a 5-point scale. The degree of inter-observers concordance was evaluated with Kappa statistics. Results The average CTDIvol for study group 1,2 and the control group was (3.00±1.41)mGy,(1.48±0.53)mGy and (5.51±2.21)mGy,respectively. Compared to that of the control group,the average CTDIvol decreased in the study group 1 and the study group 2 was 45.55% and 73.14%,while DLP and ED decreased in these two study groups (48.58% and 73.72%). The ED in study group 2 decreased to (0.71±0.28)mSv. The images of each group had acceptable image quality,and there was inter-observers agreement in diagnosis acceptability. Conclusion HDCT can be used to obtain consistent image quality with remarkable radiation dose reduction.
出处 《中国医学影像技术》 CSCD 北大核心 2010年第12期2302-2305,共4页 Chinese Journal of Medical Imaging Technology
关键词 胸部 体层摄影术 X线计算机 Thorax Tomography X-ray computed
  • 相关文献

参考文献10

  • 1Lee CI,Haims AH,Monico EP,et al.Diagnostic CT scans:assessment of patient,physician,and radiologist awareness of radiation dose and possible risks.Radiology,2004,231(2):393-398.
  • 2Hall EJ,Brenner DJ.Cancer risks from diagnostic radiology.Br J Radiol,2008,81(965):362-378.
  • 3Brenner DJ,Ellison CD.Estimated radiation risks potentially associated with full-body CT screening.Radiology,2004,232(3):735-738.
  • 4Fujii K,Aoyama T,Koyama S,et al.Comparative evaluation of organ and effective doses for pediatric patients with those for adults in chest and abdominal CT examinations.Br J Radiol,2007,80(956):657-667.
  • 5Prakash P,Kalra MK,Digumarthy SR,et al.Radiation dose reduction with chest computed tomography using adaptive statistical iterative reconstruction technique:initial experience.J Comput Assist Tomogr,2010,34(1):40-45.
  • 6Kalra MK,Rizzo S,Maher MM,et al.Chest CT performed with z-axis modulation:scanning protocol and radiation dose.Radiology,2005,237(1):303-308.
  • 7张祺丰,彭芸,李剑颖,刘玥,曾津津.64排螺旋CT自动管电流调节技术降低儿童胸部扫描剂量的应用研究[J].中国医学影像技术,2009,25(9):1591-1594. 被引量:42
  • 8Rampado O,Marchisio F,Izzo A,et al.Effective dose and image quality evaluation of an automatic CT tube current modulation system with an anthropomorphic phantom.Eur J Radiol,2009,72(1):181-187.
  • 9Prasad SR,Wittram C,Shepard JA,et al.Standard-dose and 50%-reduced-dose chest CT:comparing the effect on image quality.AJR Am J Roentgenol,2002,179(2):461-465.
  • 10Kanal KM,Stewart BK,Kolokythas O,et al.Impact of operator-selected image noise index and reconstruction slice thickness on patient radiation dose in 64-MDCT.AJR Am J Roentgenol,2007,189(1):219-225.

二级参考文献9

  • 1李剑颖.GE VCT——高图像质量和低剂量的和谐[J].CT理论与应用研究(中英文),2006,15(3):57-60. 被引量:5
  • 2Itoh S,Ikeda M,Schweden F,et al.Lung cancer screening:minimum tube current required for helical CT.Radiology,2000,215(1):175-183.
  • 3Brener DJ,Elliston CD,Hall EJ,et al.Estimated risks of radiation-induced fatal cancer from pediatric CT.AJR Am J Roentgenol,2001,176(2):289-296.
  • 4Hojreh A,Kainberger F,Puig S.Low dose multislice CT in the pediatric patient.Radiology,2003,43(12):1051-1055.
  • 5Mettler FA Jr,Wiest PW,Locken JA,et al.CT scanning:patterns of use and dose.J Radiol Prot,2000,20(4):353-359.
  • 6Greess H,Nmayr A,Wolf H,et al.Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose).Eur Radiol,2002,12(6):1571-1576.
  • 7Prasad SR,Wittram C,Shepard JA,et al.Standard-dose and 50%-reduced-dose chest CT:comparing the effect on image quality.AJR Am J Roentgenol,2002,179(2):461-465.
  • 8中島正俊.撮影条件自動適正化機能を用いたル-チン化への検討.日本放射線技術学会雑誌,2002,58(4):502-510.
  • 9Kalra MK,Rizzo S,Maher MM,et al.Chest CT performed with z-axis modulation:scanning protocol and radiation dose.Radiology,2005,237(1):303-308.

共引文献41

同被引文献119

  • 1吴晓华,马大庆,张忠嘉,冀景玲,张岩松.多层螺旋CT胸部低剂量扫描发现肺结节的临床研究[J].中华放射学杂志,2004,38(7):767-770. 被引量:86
  • 2陈胜利,黄齐好,朱栋梁,邹蓉珠,陈国东,黄子诚.冠心病介入诊治中患者的X射线辐射评估[J].中国动脉硬化杂志,2004,12(4):441-444. 被引量:6
  • 3魏宝杰,张镭,翟仁友,王亚杰.低剂量双螺旋CT薄层扫描筛查早期肺癌初步研究[J].中国医学影像技术,2003,19(11):1519-1522. 被引量:27
  • 4周海,周楠,徐霓霓,徐冰.46例隐匿性肺癌并颅内单发转移回顾性分析[J].首都医科大学学报,2006,27(2):253-254. 被引量:2
  • 5Prakash P, Kalra MK, Diqumarthy SR, et al. Radiation dose re- duction with chest computed tomography using adaptive statistical iterative reconstruction technique: initial experience. J Comput Assist Tomogr, 2010,34(1):40-45.
  • 6Singh S, Kalra MK, Gilman MD, et al. Adaptive statistical itera- tive reconstruction technique for radiation dose reduction in chest CT: a pilot study. Radiology, 2011,259(2):565-573.
  • 7European Study Group of Radiologists and physicists. European guidelines on quality criteria for computed tomography. [2009-09- 12 ] . http://www, drs. dk/guidelines/ct/quality/download/eur16262, w51.
  • 8Sone S, Takashima S, Li F, et al. Mass screening for lung cancer with mobile spiral computed tomography scanner. Lancet, 1998, 351(9111) :1242-1245.
  • 9Diederich S, Lenzen H. Radiation exposure associated with ima- ging of the chest: comparison of different radiographic and com- puted tomography techniques. Cancer, 2000,89(11 Suppl) : 2457- 2460.
  • 10Compagnonel G, Pagan L, Baleni MC, et al. Patient dose in dig- ital projection radiography. Radiat Prot Dosimetry, 2008, 129(1- 3) : 135-137.

引证文献11

二级引证文献314

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部