期刊文献+

微波消解石墨炉原子吸收光谱法测定松花粉中铅的测量不确定度评定 被引量:4

Evaluation of Measurement Uncertainty for Determination of Lead in Pine Pollen by Microwave Digestion-GFAS Method
下载PDF
导出
摘要 介绍了石墨炉原子吸收光谱法测定松花粉中铅的不确定度评定方法。通过不确定度评定得出不确定度分量的主要来源为样品制备过程所带来的不确定度,其中,消解样品所引入的不确定度贡献最大;根据最小二乘法拟合校准曲线计算的标准不确定度次之;而重复性实验以及标准储备液校准稀释所带来的不确定度相对于前两者小很多。 The evaluating method of uncertainty was introduced for determining the content of lead in pine pollen by microwave digestion - GFAS method. The results show that the key factors of uncertainty of lead content measurement and important source of indefinite component are digestion sample. evaluation indicated the uncertainty of standard curve was the second source of unccrtainty. uncertainty from standard stock solution adjusted and diluted and the repeated detcrmination The The aresubordinate factors.
出处 《广东微量元素科学》 CAS 2010年第11期55-60,共6页 Trace Elements Science
关键词 松花粉 不确定度 pine pollen lead uncertainty
  • 相关文献

参考文献4

  • 1中国合格评定国家认可委员会.CNAS-CL07测量不确定度评估和报告通用要求[S].2006.
  • 2国家技术监督局.JJG196-2006常用玻璃量器[S].北京:中国计量出版社,2006.
  • 3国家质量技术监督局.JJF1059-1999测量不确定度评定与表示[S].北京:中国计量出版社,1999..
  • 4中华人民共和国卫生部,中国国家标准化管理委员会.GB 5009.12-2010食品安全国家标准食品中铅的测定[S].北京:中国标准出版社,2010.