胸部肿瘤影像引导放疗剩余摆位误差分析 被引量:4

Analysis of Residual Set-up Errors for Imaging-guided Thoracic Tumor Radiation
下载PDF
导出
摘要 目的量化分析胸部肿瘤影像引导放疗剩余摆位误差,为制定合理的临床靶体积(CTV)到计划靶体积(PTV)外放边界提供依据。方法收集2008年3月至2008年8月,在复旦大学附属肿瘤医院行根治性放疗的胸部肿瘤患者21例,1例患者自动退出,20例进入最终研究分析。每周行一次千伏锥形束CT(KVCBCT)引导分析。以2mm作为摆位误差是否在线校正的界值,分析在线校正前、后患者的摆位误差。结果 20例患者共获取225幅KVCBCT影像。采用KVCBCT引导放疗技术校正前,患者在前后(anterior posterior,AP)、上下(superior inferior,SI)和左右(left right,LR)三个方向上的摆位误差分别为:(-0.2±4.6)mm、(0.5±4.5)mm和(-0.8±4.4)mm。根据van等[5]提供的公式:2.5∑+0.7σ,CTV到PTV的预留边界应设定为12mm。采用KVCBCT在线引导放疗技术校正后,患者在LR、SI和AP三个方向上的剩余摆位误差分别为:(-0.2±1.3)mm、(0.1±1.3)mm和(0.3±1.3)mm,CTV到PTV的预留边界应设定为3mm。结论采用KVCBCT在线引导放疗技术可以减少患者摆位误差,提高治疗精确性、保证治疗质量。 Objective To assess the residual set-up error of the on-line kilovoltage cone-beam CT (KVCBCT) guided radiotherapy for thoracic tumor patients.Methods From Mar 2008 to Aug 2008,21 patients with pathologically confirmed thoracic tumor were included in this study.On-line imaging-guided set-up error correction was performed weekly for each patient.Set-up error,before and after correction,was documented.Results A total of 225 kVCBCT scans were obtained during 6 weeks.The initial average set-up error(x-±s) was(-0.2±4.6)mm,(0.5±4.5)mm and (-0.8±4.4)mm in the left-right (LR),superior-inferior (SI) and anterior-posterior (AP) direction,respectively.After online correction,average residual set-up error (x-±s) was(-0.2±1.3)mm,(0.1±1.3)mm and (0.3±1.3)mm in the LR,SI and AP directions,respectively.According to the formalisms suggested by van Herk et al,3mm can be applied to the margin of clinical target volume (CTV) to planning target volume (PTV),when using the on-line correction.Conclusion On-line KVCBCT guidance can improve the set-up accuracy of thoracic tumor patient's radiation.
出处 《肿瘤防治研究》 CAS CSCD 北大核心 2010年第12期1430-1432,共3页 Cancer Research on Prevention and Treatment
基金 上海市科委临床医学引导项目(09411968400)
关键词 影像引导放疗 千伏锥形束CT 剩余摆位误差 Image-guided radiotherapy Kilovoltage cone-beam CT Residual set-up error
  • 相关文献

参考文献10

  • 1Dawson LA, Jaffray DA. Advances in image-guided radiation therapy [J]. J Clin Oncol, 2007,25(8):938-946.
  • 2van Herk M. Errors and margins in radiotherapy[J]. Semin Radiat Oncol, 2004, 14( 1 ) :52-64.
  • 3Letourneau D, Martinez AA, Lockman D, et al. Assessment of residual error for online cone-beam CT-guided treatment of prostate cancer patients [J]. Int J Radiat Oncol Bid Phys, 2005, 62(4) : 1239-1246.
  • 4Borst GR, Sonke J J, Bergen A. et al. Kilo voltage cone-beam computed tomography setup measurements for lung cancer patients; first clinical results and comparison with electronic portal-imaging device [J]. Int J Radiat Oncol Biol Phys, 2007, 68 (2) :555-561.
  • 5van Herk M, Remeijer P, Rasch C, et al. The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy [J]. Int J Radiat Oneol Biol Phys. 2000, 47(4):1121-1135.
  • 6戴建荣,胡逸民.图像引导放疗的实现方式[J].中华放射肿瘤学杂志,2006,15(2):132-135. 被引量:184
  • 7王艳阳,傅小龙,夏冰,吴正琴,樊旼,杨焕军,徐志勇,蒋国樑.肺癌影像引导放疗人工图像配准方法分析[J].生物医学工程学进展,2008,29(3):156-160. 被引量:9
  • 8王艳阳,傅小龙,徐志勇,徐挺,邓欣欣,蒋国梁.千伏锥形束CT引导放疗系统最小误差与剩余误差分析[J].中华放射医学与防护杂志,2008,28(2). 被引量:5
  • 9Chen YJ, Han C, Liu A. et al. Setup variations in radiotherapy ot esophageal cancer: evaluation by daily megavohage computed tomographic localization [J]. Int J Radiat Oncol Biol Phys, 2007,68(5):1537-1545.
  • 10Purdie TG-, Bissonnctte JP, Franks K, et al. Cone-beam computed tomography for on-line image guidance of lung stereotactic radiotherapy: localization, verification, and intrafraction tumor position [J]. Int J Radiat Oncol BioI Phys, 2007, 68 (1):243-252.

二级参考文献33

  • 1戴建荣,胡逸民.图像引导放疗的实现方式[J].中华放射肿瘤学杂志,2006,15(2):132-135. 被引量:184
  • 2王艳阳,傅小龙.CT在影像引导下放疗中应用的历史与现状[J].中国癌症杂志,2006,16(6):448-453. 被引量:22
  • 3[1]Dawson LA,Jaffray DA.Advances in image-guided radi-ation therapy[J].J Clin Oncol,2007,25:938-946
  • 4[2]Guckenberger M,Meyer J,Vordermark D,et al Magni-tude and clinical relevance of translational and rotational patient setup errprs:a cone-beam CT study[J].Int.J.Badiat.Oncol Biol.Phys.2006,65:934-942
  • 5[3]Langen KM,Zhang Y,Andrews RD,et al.Initial experi-ence with megavoltage (MV) CT guidance for daily pros-tate alignments[J].Int.J Radiat Oncol Biol.Phys.2005,62:1517-1524
  • 6[6]Geckenberger M,Meyer J,Wilbert J,et al Cone-beam CT based image-guidance for extracranial stereotactic ra-diotherapy of intrapulmonry humors[J].Acts.Oncol.2006,45:897-906
  • 7[7]Samson MJ,van Sornsen de Koste JR,de Boer HC,et al.An analysis of anatomic landmark mobility and setup devia-tions in radiotherapy for lung cancer[J].Int.J Radiat On-col Biol phys.1999,43:827-832
  • 8[8]Van de Steene J,Van den Houvel F,Bel A,et el.Elec-tronic portal imaging with on-line correction of setup error in thoracic irradiation:clinical evaluation[J].Int.J Radial Oncol Biol Phys.,1998,40:967-976
  • 9.调强适形放射治疗[A].胡逸民.肿瘤放射物理学[C].北京:原子能出版社,1999.538-612.
  • 10Hanley J,Debois MM,Mah D,et al.Deep inspiration breath-hold technique for lung tumors:the potential value of target immobilization and reduced lung density in dose escalation.Int J Radiat Oncol Biol Phys,1999,45:603-611.

共引文献191

同被引文献24

  • 1上官小玲,张强克,徐建国.胸部肿瘤调强放疗中摆位误差的原因与控制[J].中国医学物理学杂志,2011,28(4):2716-2717. 被引量:10
  • 2汪志,唐虹,李锐,洪浩,孔令玲.利用模拟机对493例患者摆位误差的测量[J].中国医学物理学杂志,2011,28(4):2733-2735. 被引量:4
  • 3van Herk M,Remeijer P,Rasch C,et al. The probability of correcttarget dosage ; dose population histograms for deriving treatmentmargins in radiotherapy [ J] . Int J Radiat Oncol Biol Phys,2000,47(4) :1121 -1135.
  • 4Joep C,Stroom Ben, JM Heijmen. Geometrical uncertain ties,ra-diotherapy planning margins,and the ICRU - 62 report [ J ]. Ra-diother and Oncol,2002,64( 1) :75 - 83.
  • 5van Herk M. Errors and margins in radiotherapy [ J]. Seminars inKadiat Oncol,2004,14( 1) :52 -64.
  • 6McKenzie A,van Herk M,Mijnheer B. Margins for geometric uncer-tainty around organs at risk in radiotherapy [ J]. Radiother and On-col,2002,62(3) :299-307.
  • 7戴建荣.图像引导放疗的实现方式[A].殷蔚伯,余子豪,徐国镇,等.肿瘤放射治疗学[M].北京:中国协和医科大学出版社,2008:175 -179.
  • 8DC Schiffner,TE Schultheiss,YJ Chen,et al. Can less - than - dai-ly Image guided radiation therapy (IGRT) be used to treat esopha-geal cancer? A study of patient positioning errors and their dosimet-ric consequencesf J]. Int J Radiat Oncol Biol Phys,2008,72( 1):S537 -S538.
  • 9Verhey U. Immobilizing and positioning patients for radio?therapy[J]. Semin Radiat Oncol, 1995,5(2):100-114.
  • 10van Lin EN, van der Vight L, Huizenga H, et al. Set-up improvement in head and neck radiotherapy using a 3D off-line EPID-based correction protocol and a customised head and neck support[J]. Radiother Oncol, 2003, 68(2): 137-148.

引证文献4

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部