期刊文献+

实验室内多台血细胞分析仪的校准和比对分析 被引量:11

Calibration and comparative analysis among different hematology analyzers
下载PDF
导出
摘要 目的:通过对不同血细胞分析仪的测定结果进行比对,探讨不同的血细胞分析仪检测结果之间的差异是否在允许范围之内,以提高检验结果的准确性、一致性。方法:以一台性能较好的血细胞分析仪作为标准仪器,用新鲜全血进行同一实验室内多台血细胞分析仪的比对试验,检测结果参照SYSMEX公司的仪器说明书提供的标准作为各参数的允许范围进行评估。结果:除XS-800i的WBC超出允许范围外,其他比对仪器检测结果均在偏差允许范围之内,检测结果准确性、一致性良好。结论:实验室内拥有多台血细胞分析仪时,应通过定期比对试验及定值新鲜全血校准比对仪器,建立多台血细胞分析仪的校准和质量控制方法,保证实验室检测结果的准确性和一致性。 Objective To improve the accuracy and consistency,we compared the test results from the different blood cell analyzers to explore whether the variance of the results were within the permitted range.Method The one hematology analyzer with best performance was chosen as reference analyzer,anticoagulative fresh whole blood was detected on several hematology analyzers by verification test,then we evaluated test results with the reference provided by SYSMEX company.Results The test results are all in reference limits with good accuracy and consistency except one WBC value of XS-800i.Conclusion When the laboratory has more than one blood cell analyzer,we should regularly perform the verifying test and the calibrating with fresh whole blood to establish calibration and quality control methods of the multiple blood cell analyzers in order to ensure the accuracy of laboratory test results and consistency.
出处 《吉林医学》 CAS 2011年第1期19-20,共2页 Jilin Medical Journal
关键词 血细胞分析仪 校准 比对试验 Hematology analyzer Calibration Verification test
  • 相关文献

参考文献4

二级参考文献10

  • 1宋颖,胡晓波,居漪,李泳,许蕾,朱秀兰,金大鸣.红细胞计数参考方法的不确定度研究[J].检验医学,2005,20(3):185-188. 被引量:7
  • 2ASTM F 2149-01.Standard tex method for automated analyses of cell-the electrical sensing zone method of enumerating & Sizing cell suspension[S].2002.
  • 3International Council for Standardization in Haematology.Reference method for the enumeration of erythrocytes and leukocytes[J].Clin Lab Haemat,1994,16(2):131-138.
  • 4ICSH.Guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting and cell marker applications[J].Clin Lab Haemat,1994,16 (2):157-174.
  • 5Lewis SM.Harmonization and standardization of the blood countthe role of the ICSH cytometry panel-[].Sysmex J Int.1999
  • 6Bain,Barbara J.Detecting erroneous blood counts, chapter 4 (132146) of blood cells -a practical guide- 2nd Edition, Blackwell Science[].United Kingdom.1995
  • 7Cornbleet J.Spurious results from automated hematology cell counters[].Laboratory Medicine.1983
  • 8ICSH HGB Expert Panel.Recommendation for reference method for haemoglobinometry in human blood and specifications for international haemiglobincyanide standard[].Journal of Clinical Pathology.1996
  • 9ICSH Cytometry Panel.Platelet counting by the RBC/patelet Ratio Method-A reference Method-[].American Journal of Clinical Pathology.2001
  • 10ICSH Cytometry Panel.Recommendation for Reference method for the packed cell volume[].Laboratory Hematology.2001