摘要
目的 研究3种无源固体探测器对低能X射线的能量响应,评估在乳腺X射线摄影检查中测量入射体表剂量和估算平均腺体剂量的可行性.方法 选用LiF(Mg,Ti)和LiF(Mg,Cu,P)热释光(TL)材料和Al2O3:C光致发光(OSL)材料的固体探测器,在Mo/Mo低能X射线机上,管电压25~40 kV范围的10种线质分别曝光相同的空气比释动能.结果 给出管电压在25 kV~40 kV范围的10种辐射线质时TL和OSL能量响应,分别小于25%和11%.结论 Al2O3:C光释光探测器用于乳腺X射线摄影日常质量控制和入射体表剂量测量具有相当大的潜力.
Objective To assess the feasibility of ESD measurements in the mammography by the comparative study of energy dependence of the three passive solid detectors on low energy X-ray.Methods Two thermoluminescent TL) detectors ( LiF(Mg,Ti) and LiF( Mg,Cu, P) and one optically stimulated luminescence (OSL) detector( Al2O3: C ) were irradiated for the same air kerma by ten kinds of beam quality between 25-40 kV in a Mo/Mo low energy X-ray generator,respectively. Results The energy dependences of TL and OSL were 25% and 11% between 25-40 kV, respectively. Conclusions considerable potential for OSL detector can be used in routine quality control and ESD measurements in the mammography.