功能磁共振的质量控制方案分析 被引量:3

Analysis on DQA Protocol of fMRI
原文传递
导出
摘要 本文通过对水模扫描,获取数据质量控制(DQA)参数,以功能磁共振(fMRI)序列获取序列图像进行信噪比、漂移等计算,漂移计算方法简化为对兴趣区(ROI)区域平均信号强度的极值处理。长期应用该方法的实践表明,该方法能有效地反映出设备的成像性能与稳定性,是一种简便可行的fMRI质量控制方法。 Our purpose is to introduce and analyze the data quality assurance(DQA) protocol of functional magnetic resonance imaging(fMRI).A water phantom was scanned to get DQA indexes.An fMRI sequence was used to get signal noise ratio(SNR)and Drift,which was calculated from maximum difference ratio of the average signal intensity in the region of interest(ROI) of image serials.The long period application of this method demonstrated that this DQA protocol can reflect imaging performance and the state of stability of the MRI scanner.Some application experience and discussion involved in DQA were also presented here.
出处 《生物医学工程学杂志》 EI CAS CSCD 北大核心 2010年第6期1247-1250,共4页 Journal of Biomedical Engineering
关键词 功能磁共振成像 数据质量控制 信噪比 漂移 Functional MRI Data quality assurance(DQA) Signal noise ratio(SNR) Drift
  • 相关文献

参考文献1

二级参考文献6

  • 1Ronald RP, Jerry A, Richard JM, et al. Pratical aspects of functional MRI[J]. Med Phys, 2002, 29(8): 1892-912.
  • 2Moonen CT, Bandettini PA. Functional MRI[M]. Berlin: SpringerVerlag. 337-8.
  • 3WeisskoffRM. Simple Measurement of Scanner Stability for Functional NMR Imaging of Activation in the Brain[J]. MRM, 1996, 36:643-5.
  • 4陈平雁.SPSS10·0统计软件应用教程[M].北京:人民军医出版社,2001.274-5.
  • 5Ogawa S, Tank DW, Menon R, et al. Intrinsic signal changes accompanying sensory stimulation: functional brain mapping with magnetic resonance imaging[J]. Proc Natl Acad Sci USA, 1992, 89: 5951-5.
  • 6Thulbom KR, Waterton JC, Matthews PM, et al. Dependence of the transverse relaxation time of water protons in whole blood at high field[J]. Biochim BiophysActa, 1982, 714: 265-0.

共引文献5