X射线发生器实际峰值电压的测量

Practical peak voltage of X-ray generator
下载PDF
导出
摘要 目的初步分析X射线发生器实际峰值电压(PPV)的测量误差及其测量差异。方法采用PPV测量方法对不同X射线成像设备进行测量。结果平均峰值电压的测量误差为-1.4~+1.7kV,最大峰值电压测量变化为-0.9~+2.2kV,PPV测量偏差为-3.7~-0.8kV。结论由于计算方法的不同,PPV测量值普遍低于设定电压值和平均值,因而总体测量误差高于平均值和最大峰值电压的测量误差。 Objective To analyze the measurement variation for practical peak voltage(PPV).Methods Different X-ray imaging equipments were measured by PPV method.Results The variation of average peak voltage was-1.4 to + 1.7 kV and the maximum peak voltage measurement was-0.9 to + 2.2 kV and the actual peak voltage measurement error was-3.7 to-0.8 kV.Conclusion Since the calculation methods are different,the PPV values are generally lower than kV settings and average peak voltage,thus the overall measurement error of PPV is higher than the measurement error of average and maximum peak voltage.
作者 白玫 严汉民
出处 《生物医学工程与临床》 CAS 2011年第1期82-85,共4页 Biomedical Engineering and Clinical Medicine
关键词 X射线发生器 实际峰值电压 平均峰值电压 X-ray generator practical peak voltage mean peak voltage
  • 相关文献

参考文献8

  • 1Beutel J, Kundel HL Van Metter RL. Handbook of medical ima- gint~d]. Washington, USA: SPIE Press, 2000.
  • 2Kramer HM, Selbach HJ, Iles WJ. The practical peak voltage of diagnostic X-ray generators[J]. BrJ Radiol, 1998, 77(842): 200- 209.
  • 3Ranallo FN. Doctor thesis: the noninvasive measurement of X- ray tube potential[D]. Madison: University of Wisconsin, 1993.
  • 4International Electrotechnical Commission. IEC 61676: Medic- al electrical equipment- dosimetric instruments used for non- invasive measurement of X-ray tube voltage in diagnostic radiology[M]. International Electrotechnical Commission, Gene-va, Switzerland, 2002.
  • 5Baorong Y, Kramer HM, Selbach HJ, et al. Experimental deter- mination of practical peak voltage[J]. Br J Radiol, 2000, 73(870): 641--649.
  • 6None. TINT 12000 X-ray test device users manual PN 132000- 1001[M]. Fluke Biomedical Corporation, 2008.
  • 7Terini RA, Potiens M, Herdade SB, et al. Measurement of the qu- antity practical peak voltage in the radiology practice[J]. Radiol Bras, 2009, 42(6):389-394.
  • 8Kramer HM, Selbach HJ. Extension of the range of definition of the practical peak voltage up to 300 kV[J]. BrJ Radiol, 2008, 81 (969): 693-701.

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部