期刊文献+

上槽泄露导致双向电泳凝胶高分子量区域垂直拖尾

Leaking of Buffer from Upper Chamber Caused Vertical Streaking in the High Molecular Weight Separation with 2-DE Gels
下载PDF
导出
摘要 本文探讨了电泳系统上槽泄露引起双向电泳凝胶蛋白点纵向拖尾问题.上槽泄露导致上槽SDS损耗,电流增大以及局部温度升高从而导致双向电泳凝胶高分子量区域蛋白点纵向拖尾.上槽泄露与上槽设计存在不足,使用者不适当操作以及上槽维护不良有关.对Ettan DALT six电泳系统而言,一旦出现上槽泄漏问题,建议更换改良后的新版上槽.电泳过程中应避免上槽泄露,SDS的不足或损耗,电流过大和电泳液温度过高. This paper intended to solve the problem of vertical streaking caused by the leaking of upper buffer chamber(UBC) in two-dimensional gel electrophoresis(2-DE).UBC leakage resulted in sodium dodecyl sulfate(SDS) depletion,increased current and elevasted local temperature of the upper buffer,leading to vertical streaking in the high-molecular-weight region of 2-DE gels.The leaking problems could be caused by either the defected design of the 2-DE system or the improper usage /maintenance during operation.An improved Ettan UBC is recommended if the leaking is observed with an existing system to avoid the causes of unsatisfactory separation of electrophoresis.Meanwhile,the operation and maintenance following the manufacturers instructions is also very important.
出处 《中国生物化学与分子生物学报》 CAS CSCD 北大核心 2011年第1期96-100,共5页 Chinese Journal of Biochemistry and Molecular Biology
基金 国家自然科学基金项目(No.20637010 30570420)~~
关键词 双向电泳 垂直拖尾 上槽泄露 two-dimensional gel electrophoresis vertical streaking leaking of the upper buffer chamber
  • 相关文献

参考文献10

  • 1Wittmann-Liebold B, Graack H R, Pohl T. Two-dimensional gel electrophoresis as tool for proteomics studies in combination with protein identification by mass spectrometry [ J ]. Proteomics, 2006, 6( 17 ) :4688-4703.
  • 2Berkelman T, Brubacher M G, Chang H, et al. Tips to prevent streaking on 2-D gels. Bio-Rad Bulletin3310, 2004, gemonicproteomictechnology, com/articles/abl/bl 105 bet. pdf.
  • 3Westermeier R, Naven T, Hspker H R. Proteomics in Practice A guide to successful experimental design [ M ]. Weinheim Wiley-VCH Verlag GmbH & Co. KGaA, 2nd Ed 2008.
  • 4GE Healthcare. Ettan DALT six electrophoresis system user manual. GE Healthcare, 2006, Uppsala, Sweden.
  • 5GE Heahhcare. 2-D eleetrophoresis-prineiples and methods. GE Heahhcare, 2004, Uppsala, Sweden.
  • 6Gorg A, Weiss W, Dunn M J. Current two-dimensional electrophoresis technology for protenmies [ J ]. Proteomies, 2004, 4( 12 ) :3665-3685.
  • 7Berkelman T, Brubaeher M G, Chang H. Prevention of vertical streaking[ J ]. BioRadiations, 2004, 112:23-24.
  • 8Sehagger H. Tficine - SDS-PAGE [ J]. Nature Protoc, 2006, 1 ( 1 ) :6-22.
  • 9Andrea B L. Dupont. Role of NOS-like proteins found in bacteria [ D]. University of Waterloo, Canada 2006.
  • 10Starita-Geribaldi M, Poggioli S, Zucchini M, et al. Mapping of seminal plasma proteins by two-dimensional gel electrophoresis in men with normal and impaired spermatogenesis [ J ]. Mol Hum Reprod, 2001, 7(8) :715-722.