将改进的国产WLK-6A型阴离子抑制器应用于Dionex 600型离子色谱系统,对基线噪声、基线漂移、重复性及实际样品测定中的线性、检出限等作了试验。结果表明,改进的WLK-6A型抑制器可以与戴安AS14A(3mm)、AS14A(4mm)柱子配套使用。
The improved domestic WLK-6A anion suppressor was applied to Dionex ion chromatographic system. The baseline noise, baseline drift and reproducibility were tested. The linearity and detection limit for actual sample were investigated. The results show that the improved WLK-6A anion suppressor is suitable to be used with Dionex AS14A 3mm and Dionex AS14A 4mm columns.
出处
《分析仪器》
CAS
2011年第1期78-82,共5页
Analytical Instrumentation