基于新颖血压参数标定法的无袖套血压获取系统

Cuffless Blood Pressure Acquisition System Based on a Novel Calibration Method
下载PDF
导出
摘要 基于Moens-Korteweg模型和流体静力学原理,提出了一种简易人体血压参数标定方法,通过改变手腕血压测量点与心脏的垂直距离,用流体静力压来影响血液在血管中的流动,改变血压与脉搏波传导时间,从而得到多组不同的心血管参数值,由此建立起血压与脉搏波传导时间的关系式。文中详细介绍了各部分电路及软件设计。通过对不同年龄、性别的人分别进行长期和短期的监测,并与医用OMRON EW3152电子血压计进行对比,结果表明用此方法测量有较高的精确度和测试一致性。 Based on Moens-Korteweg model & hydrostatic principle, a simple calibration method by changing vertical distance between heart and radial artery is proposed in this paper. Using hydrostatic pressure to change arterial pressure and pulse wave transmit time, consequently obtain several sets of cardiovascular parameters and finally build up relationship between blood pressure and pulse wave transmit time. Clinical tests have been carried with different age and gender for long and short time monitoring. The comparison results with Sphygmomanometer OMRON EW3152 show its promising accuracy and coincidence feature in blood pressure measurement.
出处 《中国医疗器械杂志》 CAS 2011年第1期6-10,共5页 Chinese Journal of Medical Instrumentation
基金 澳门FDCT资助 编号:018/2009/A1
关键词 简易血压参数标定 流体静力学 脉搏波传导时间 动态无袖套血压测量 simple blood pressure calibration method, hydrostatic principle, pulse wave transmit time, cuffless, ambulatory blood pressure measurement
  • 相关文献

参考文献1

二级参考文献10

  • 1陆渭明,李和平,蒋正尧,崔鲁宁,陶上乘,张冬梅,涂继善,缑东元.对利用脉搏波传播速度间接测量血压时主要影响因素的研究[J].中国生物医学工程学报,1995,14(1):89-93. 被引量:14
  • 2朱雯,阮晓声,梁中庆,薛大建.脉动波在弹性管道中的传播速度与诸因素的关系[J].浙江医科大学学报,1996,25(4):154-156. 被引量:8
  • 3Beate H.McGhee.Monitoring arterial blood pressure:what you may not know[J].Critical care Nurse,2002,22(4):60~78
  • 4Pressman,G.L.,P.M.Newgard.A transducer for the continuous external measurement of arterial blood pressure[J].IEEE Trans Biomed Electron,1963,10:73~81
  • 5Tanaka S,Yamakoshi K.Ambulatory instrument for monitoring indirect beat-to-beat blood pressure in superficial temporal artery using volume-compensation method[J].Med Biol Eng Comput,1996,34(6):441~447
  • 6Bazzett HC,Dfeyer NB.Measurement of pulse wave velocity[J].Am.J.Physiol,1922,63:94~115.
  • 7Chen W,Kobayashi T,Ichikawa S,et al.Continuous estimation of systolic blood pressure using the pulse arrival time and intermittent calibration[J].Med Biol Eng Comput,2000,38(5):569~574
  • 8Pitson DJ,Stradling JR.Value of beat-to-beat blood pressure changes,detected by pulse transit time,in the management of the obstructive sleep apnoea/hypopnoea syndrome[J].Eur Respir J,1998,12(3):685~692
  • 9Lansdown M.A method of using induced waves to study pressure propagation in human arteries[J].Circ.Res,1957,5:594~601
  • 10俞梦孙,吕沙里,张春艳,等.一种利用脉搏波测量动脉血压的方法与装置.中国专利.ZL95108190.X,1997,2

共引文献30

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部