用两种人体胸腔模型对超铀元素活体测量系统刻度的结果及其比较

CALIBRATION OF IN VIVO MEASUREMENT SYSTEM FOR TRANSURANIC ELEMENTS AND RESULTS COMPARISON USING LLNL AND JAERI PHANTOMS
下载PDF
导出
摘要 本文报道用LLNL胸腔模型和JAERI躯干模型对由双层晶体探测器构成的超铀元素活体测量系统进行效率刻度的结果及其比较。实验结果表明,用JAERI模型得到的肺中241Am的计数效率比LLNL模型的高7%~10%;而对肺中238,239Pu用LLNL模型得到的计数效率比用JAERI模型的高7%~20%。 This paper reports some experimental results of efficiency calibration and comparison between them using LLNL and JAERI torso phantoms for in vivo monitoring system involving a phoswich detector for measurement of transuranic elements. The experimental results show that the lung counting efficiencies obtainted using JAERI torso phantom are 7% ̄10% higher than that obtainted using LLNL phantom for 241 Am lung sets, and for 238, 239 Pu lung sets, the lung counting efficiencies from LLNL phantom are 7% ̄20% higher than that from JAERI phantom. (
出处 《辐射防护》 CAS CSCD 北大核心 1999年第4期263-271,共9页 Radiation Protection
关键词 超铀元素 活体测量 体模 刻度 肌肉等效胸壁厚度 Transuranium,In Vivo Measurement,Phantom,Muscle Equivalent Chest Wall Thickness)