摘要
经检验,从过氧乙酸消毒的复用透析器采样液中未检出细菌。临床观察10000例次使用患者,未发现交叉感染。使用中血透并发症发生率为008%,远低于用福尔马林消毒者(105%)。在5例肾囊肿病例中未发现癌变(用福尔马林消毒者4例中有2例癌变),复用的透析器对尿素氮及肌酐清除率均达正常标准。
Examination did not detect bacteria in sample fluid taken from reused dialyzator disinfected with peracetic acid. No cross infection was found in clinical observation of 10 000 case-occasions using this dialyzator. The incidence of complications of hemodialysis during use of the dialyzator was 008% which was far lower than that in formalin disinfection (105%). No cancerous change was found in 5 cases of renal cysts (2 cases of cancerous change were found in 4 cases using formalin disinfection). The urea nitrogen and creatinine clearance of the reused dialyzator attained normal standard.
出处
《中国消毒学杂志》
CAS
1999年第2期100-101,共2页
Chinese Journal of Disinfection