摘要 双柱双检测器气相色谱法同时进样分析非甲烷烃,对甲烷、总烃、氧气和仪器引入的不确定度进行了评定。结果显示,本法测定不确定度的主要来源为标准气体的不确定度,其次是色谱分析仪器引入的不确定度,相对分子质量引入的不确定度可忽略不计。环境空气中非甲烷烃测定结果为(0.85±0.14)mg/m3,k=2。 Non-methane hydrocarbons in the gas sample were analyzed simultaneously by GC with double columns and double FIDs. The uncertainty components of methane, hydrocarbons, oxygen and instruments were evaluated. The results showed that the main source of the uncertainty was the uncertainty of standard gas, the second was the gas chromatography, and the uncertainty of mo- lecular weight could be ignored. So the detection results of non-methane hydrocarbons was (0.85 ±0.14) mg/m3 ,k = 2.
出处 《化学分析计量》 CAS 2011年第1期15-17,共3页 Chemical Analysis And Meterage
关键词 不确定度 非甲烷烃 双柱双检测器 同时进样 uncertainty, non-methane hydrocarbon, double column and double FID, simultaneous sampling
  • 相关文献

参考文献6

二级参考文献16

  • 1李定邦,陈浩,张大年.直接法测定大气中的非甲烷烃[J].上海环境科学,1995,14(6):26-28. 被引量:9
  • 2孙东晓,张大年.大气中非甲烷烃的测定研究进展[J].上海环境科学,1996,15(4):11-15. 被引量:6
  • 3空气和废气检测分析方法(第4版)[M].北京:中国环境科学出版社.
  • 4HJ/T38-1999.固定污染源排气中非甲烷总烃的测定气相色谱法[S].
  • 5V. Mugica, E. Vega. Speciated non- methane organic compounds emissions from food cooking in Mexico [ J ]. Atmospheric Environment 35(2001 ) :1 - 5.
  • 6JJF1033-2008.计量标准考核规范[s].国家质量技术监督局.
  • 7JJF1059-1999.测量不确定度评定与表示[S].[S].,..
  • 8JJG700-1999气相色谱仪检定规程
  • 9日本化学会.碳氢化合物污染及其对策[M].北京:科学技术出版社,1987.
  • 10JJF1059-1999测量不确定度的评定与表示.