SYSMEX SE-9000血细胞分析仪2例特殊计数情况分析

下载PDF
导出
摘要 目的对SE-9000血细胞分析仪遇到的2例特殊情况进行分析,提出建立适合自己实验室的复片规则,并不断完善。方法采用SYSMEX SE-9000血细胞分析仪对血细胞进行计数。结果病例1:仪器误报细胞质内有中毒颗粒的中性粒细胞为嗜碱性粒细胞;病例2:仪器中性粒细胞和淋巴细胞计数错误。结论五分类血细胞分析仪作为技术先进、有很多特性的大型仪器,为临床工作带来了很大的方便,但也有其不足之处,实验室应根据自身的特点,结合仪器状态和患者情况,特别是对没有异常信息提示的标本,建立符合自身特点的复检规则。
作者 陈敏 曹军皓
出处 《检验医学与临床》 CAS 2011年第4期461-461,472,共2页 Laboratory Medicine and Clinic
  • 相关文献

参考文献9

二级参考文献35

  • 1马洪玉,朱忠勇.遗传性中性粒细胞过氧化物酶缺乏[J].国外医学(临床生物化学与检验学分册),2005,26(4):255-255. 被引量:4
  • 2International Council for Standardization in Haematology. Guidelines for the evaluation of blood cell analyzer including those used for differential leukocyte and reticulocyte counting and cell marker application. Prepared by the ICSH expert panel on cytometry [ J ].Clin Lab Haemat, 1994,16:157-174.
  • 3National Committee for Clinical Laboratory Standards. Leukocyte Differential Counting NCCLS Approved Standard H20-A2 [ S ].Villanova, PA, 1992.
  • 4Harrison P, Horton A, Grant D, et al. Immunoplatelet counting: a proposed new reference procedure [ J ]. Br J Haematol, 2000, 108:228-235.
  • 5Hallawell R,O Malley C,Hussein S,et al. An evaluation of the Sysmex NE-8000 hematology analyzer[J]. Am J Clin Pathol ,1991,96 :594-601.
  • 6Carter A B, Chavda N, Grant D,et al. Evaluation of the Sysmex SE-9000 automated hematology analyzer[ J]. Sysmex J Int, 1995,5:6-16.
  • 7Vives-Corrons J L,Besson I,Jou J M,et al. Evalution of the Abbott Cell-DYN 3500 hematology analyzer in a university hospital[ J]. Am J Clin Pathol, 1996,105:553-559.
  • 8Thalhammer-Seherrer R, Knobl P, Korninger L, et al. Automated fivepart white blood cell differential counts, efficiency of softwaregenerated white blood cell suspect flags of the hematology analyzers Sysmex SE-9000, Sysmex NE-8000, and Coulter STKS [ J ]. Arch Pathol Lab Med, 1997,121:573-577.
  • 9Bentley S A, Johnson A, Bishop C A. A parallel evaluation of four automated hematology analyzers [ J ]. Am J Clin Pathol, 1993,100:626-632.
  • 10Barnes PW, McFadden SL, Machin SJ,et al. The International Group for Hematology Review:suggested crite- ria for action following automated CBC and WBC diferential analysis [ J ]. Lab Hematol, 2005,11 ( 1 ) : 83- 90.