期刊文献+

U型环固定法在医疗器械清洗中的应用研究 被引量:2

APPLICATION OF U-RING FIXATION TO THE CLEANING OF MEDICAL DEVICES
原文传递
导出
摘要 目的研究U型环固定法在医疗器械清洗中的应用价值,以提高手术器械上残留血迹清洗效果。方法采用胶体金试纸法,对应用U型环固定法固定器械的清洗效果进行了评价,同时与未用U型环固定法的对照组作平行比较。结果将器械轴节打开以双向串在两个U型环器械搁物架上,置入全自动清洗消毒器中清洗,对器械上血迹清洗合格率达到99.50%。未采用U型环固定法的对照组器械上血迹清洗合格率为94.5%。结论使用U型环器械搁物架清洗器械明显提高了清洗效果。 Objective To study the application value of U-ring fixation in the cleaning of medical instruments in order to enhance the cleaning effect on the residual blood.Methods Colloidal gold test paper method was used to evaluate the cleaning effect of U-ring fixation on the fixed equipment,which was compared with the control group that without U-ring fixation on the same time.Results Instruments was stringed in the two U-ring object shelves with the instrument axis sections opening to two-way.Then these instruments were placed in the automatic cleaning sterilizer to be cleaned and the qualified rate of blood cleaning on them achieved 99.50%.While the qualified rate of blood cleaning on the instruments without U-ring fixation was 94.5%.Conclusion Using the U-ring object shelves to clean medical instruments can significantly improve the cleaning effect.
出处 《中国消毒学杂志》 CAS 2011年第1期21-22,24,共3页 Chinese Journal of Disinfection
关键词 医疗器械 U型环 清洗效果 试纸法 medical devices U-ring cleaning effect test paper method