摘要 目的利用Westgard标准决定图分析常用临床生化检测系统性能的可接受性。方法分别以室内质控数据计算各检测项目的日间不精密度,以室间质控回报结果计算不准确度,在Westgard标准决定图上确定标准化操作点,判断生化检测各项目的方法性能。结果参与室间质评的19个常规生化项目中,氯(Cl-),γ-谷氨酰转移酶(GGT)为临界分析性能,其余各项分析性能优良。结论通过定期以Westgard标准决定图对检测系统进行评价,可有效地发现检测系统性能的改变,为提高实验室的工作质量提供保障。 Objective To evaluate the analytical quality of usual laboratory test items with the application of Westgard method decision chart.Methods The intraday inaccuracy was calculated from external quality control data and the imprecision from the internal quality control data.The acceptability of the 19 tests was demonstrated with method decision chart.Results The performances of two items(Cl-,GGT)were marginal and all the others were perfect.Conclusion Regular decisions on the assay system with the method evaluation decision chart should be made,which would be helpful in finding the problems of the assay system effectively and providing guarantee for increasing the working quality of laboratory.
出处 《检验医学与临床》 CAS 2011年第5期541-542,共2页 Laboratory Medicine and Clinic
关键词 标准决定图 检测系统 精密度 准确度 method decision chart assay system precision accuracy
  • 相关文献

参考文献5

二级参考文献14

  • 1刘小娟,黄亨建,李萍,徐克和,余霆,胥劲,叶艾竹,吴士吉.罗氏生化试剂用于奥林巴斯全自动生化分析仪的溯源性分析[J].华西医学,2004,19(3):401-402. 被引量:16
  • 2李萍,刘小娟,徐克和,黄志刚.利用Westgard标准决定图判定测定方法性能[J].临床检验杂志,2006,24(1):69-70. 被引量:27
  • 3张秀明,庄俊华,徐宁,黄宪章,陈茶,尹一兵,梁伟雄.不同检测系统血清酶测定结果的偏倚评估与可比性研究[J].中华检验医学杂志,2006,29(4):346-349. 被引量:137
  • 4[1]Mmedicare,Medicaid and CLIA programs:regulations implementing the Clinical Laboratory Improvement Amendments of 1988(CLIA).
  • 5[2]Westgard JO.A method evaluation decision chart(MEDx chart)for judging method performance[J].Clin Lab Sci,1995,8(5):277-203.
  • 6[3]Mansouri S,Fallon K.Enhancing error detection in GEM Premier 3000 bv failure pattern recognition Presented at the AACC critical care and point of care testing conference[C].September 11-14,2002,Monterey,Califomia.
  • 7[4]Ehnneyer S,Laessig R,Fallon K.A new quality management paradigm for POCT:manu-facturer-assured quality with the GEM Premier 3000 and iQM Presented at the 15th IFCC-FESCC European Congress of Clinical Chemistry and Laboratory Medicine[C].June 1-5,2003,Barcelona,Spain.
  • 8Westgard JO.A method evaluation decision chart (MEDx Chart)for judging method performance.Clin Lab Sci,1995,8:277-283.
  • 9Ricos C,Alvarez V,Cava F,et al.Current databases on biological variation:pros,cons and progress.Scand J Clin Lab Invest,1999,59:491-500.
  • 10Westgard JO,Burnett RW.Precision requirements for costeffective operation of analytical processes.Clin Chem,1990,36:1629-1632.

共引文献276

同被引文献4

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部