期刊文献+

流动注射仪测定水中硫化物方法探讨 被引量:6

Study on Detection of Sulfide in the Water by Flow Injection Analysis Instrument
下载PDF
导出
摘要 采用德国布朗卢比AA3型流动注射分析仪测定水中硫化物,通过在线处理释放出H2S气体,在酸性介质中与对氨基二甲基苯胺和三氯化铁反应,生成亚甲蓝染料,于660nm处比色测定。方法检出限为0.004mg/L,标准曲线有很好的稳定性,灵敏度大大提高,精密度和准确度较好,操作简便易行,能满足大批量水样以及应急监测及时准确分析的需求。 The flow injection analysis instrument from Bran Luebbe of German was used to monitor the sulfide in the water. The H2S gas was released by on -line treatment to react with ferric chloride and p -aminodimethyl aniline to methylthionine chloride which was determined by colorimetric method under 660 nm. The detection limit of this method is 0. 004mg/L with stable standard curve and highly improved sensitiveness and excellent accuracy and simple operation which can meet the monitoring requirements of large number of water samples and emergent monitoring.
出处 《环境科学导刊》 2011年第2期97-100,共4页 Environmental Science Survey
关键词 流动注射仪 在线分析 水中硫化物 flow injection analysis instrument analysis on line sulfide in the water
  • 相关文献

参考文献5

  • 1本书编委会.水和废水监测分析方法[M].北京:中国环境出版社,2002.
  • 2GB/T16489-1996,地表水环境质量标准[S].
  • 3本书编委会.环境水质监测质量保证手册[M].北京:化学工业出版社,1984.
  • 4中华人民共和国水利部.水环境监测规范[S].1998.
  • 5美国环保局.水和废水监测标准方法[S].EPA-600/R-93-100.

共引文献4

同被引文献53

引证文献6

二级引证文献24

相关作者

内容加载中请稍等...