期刊文献+

垂体MR图像质量控制及扫描方式的研究

Quality Control and Scanning Method of Pituitary MR Image
下载PDF
导出
摘要 目的:探讨垂体MR检查的理想扫描方法,以提高垂体MRI图像质量。方法:选择40例疑似垂体微腺瘤志愿者的MR扫描图像,按扫描方不同分组:常规组SE序列分为A组(不加饱和技术)和B组(加饱和技术),研究组FE序列分为A1组(不加饱和技术)和B1组(加饱和技术),观察垂体及周围解剖结构和细节的显示情况,将质量分为优、良、差3级。采用SPSS统计软件对所得的数据进行统计分析。结果:FE序列加饱和技术能够清晰显示垂体及相关解剖结构。B1组的扫描方法与常规组及A1组相比,图像质量明显提高,P<0.05。结论:使用FE加饱和技术能够提高图像质量,缩短工作时间,提高工作效率。 Objective To explore the ideal pituitary MR examination methods in order to improve the pituitary MRI scan image quality.Methods 40 patients with suspected anterior pituitary adenomas were selected as volunteers for MR scanning images,which was divided into four groups,including A group without saturation technology and B group with saturation technology of SE sequence,and A1 group without saturation technology and A2 group with saturation technology of FE sequence.The anatomical structure and detail of the pituitary and the around tissue were displayed and evaluated.The display quality was divided into three levels: excellent,good and bad by SPSS for data analysis.Results FE sequence with saturation technology could show the pituitary gland and relevant anatomy clearly.Scanning method of B1 group was compared with conventional group and A1 group.The quality of image was improved obviously,with p0.05.Conclusion Using FE and saturation technology can improve the quality of image,shorten working hours,and improve work efficiency.
作者 彭俊玲
出处 《医疗卫生装备》 CAS 2011年第2期60-61,共2页 Chinese Medical Equipment Journal
关键词 垂体 磁共振成像 扫描方式 pituitary magnetic resonance imaging(MRI) scanning mode
  • 相关文献

参考文献8

二级参考文献47

  • 1陈敏,杜安涛,王占立,李果珍.腹部磁共振成像的临床应用──探讨和比较几种扫描脉冲序列[J].中华放射学杂志,1994,28(8):517-521. 被引量:4
  • 2林锐,吴仁华,肖壮伟,孔抗美,郎志谨.零点充填技术提高增强磁共振血管成像内径测量准确性的模型研究[J].医学影像学杂志,2005,15(7):596-598. 被引量:1
  • 3葛玉林.梯度回波技术与应用[J].国外医学(临床放射学分册),1996,19(2):83-89. 被引量:10
  • 4燕树林 曾祥阶.医学影像技术全科纲要及考题解,第1版[M].武汉:湖北科学技术出版社,1999..
  • 5刘定西 于群.MR成像分册,第1版[M].武汉:湖北科学技术出版社,1999..
  • 6彭振军.医用磁共振成像技术,第1版[M].武汉:湖北科学技术出版社,1998.85.
  • 7毛希安.现代核磁共振实用技术及应用,第1版[M].北京:科学技术文献出版社,1999.156.
  • 8祁吉.临床磁共振血管成像,第1版[M].北京:人民卫生出版社,1998.85.
  • 9[1]Debiao Li,Rober TW,Weili Li. Magnetic resonance imaging of the brain with Gd-DTPA: comparison of T1-weighted spin-echo and 3D gradient-echo seq uences[J]. JMRI,1996,6(3)∶415-423.
  • 10Covel M M ,et al.Automated analysis of multiple performance characteristics in magnetic resonance imaging systems[J].Med Phys,1986,13:815- 823.

共引文献51