摘要 目的:了解再生医疗器械清洗质量,提高灭菌保证水平.方法:用目测法和镜检方法检测比较除锈处理对清洗效果的影响.结果:经过除锈处理的再生医疗器械,用多酶清洗剂经超声清洗、煮沸润滑、清洗、干燥等处理步骤,目测检测符合洁净质量要求达到99%;用放大镜检测发现绣渍器械5件.未经过除锈处理的再生医疗器械直接按相同程序进行清洗,目测检查符合清洗洁净质量要求仅为84.2%;用放大镜检测发现绣蚀、锈斑的器械79件.结论:再生医疗器械必须经过除锈处理,方可确保清洗质量. Objective To understand the quality of renewable clean medical equipment to improve the sterilization assurance level. Methods to detect visual comparison method and microscopy rust on cleaning effect. Results After rust treatment regeneration medical equipment, multi--enzyme detergent used by the ultrasonic cleaning, boiling lubrication, cleaning, drying and other processing steps, visually tested and found clean 99% quality requirements; embroidered stains detected with a magnifying glass device 5. Rust has not been dealt with directly by the same regeneration process of medical devices for cleaning, washing clean visual inspection meet the quality requirements of only 84.2%; detected with a magnifying glass embroidery corrosion, pitting the device 79. Conclusion regenerative medical devices must be rust treatment, only to ensure the eleaning quality.
作者 杨小娟
出处 《医学信息(下旬刊)》 2011年第3期315-315,共1页 Medical information
关键词 再生医疗器械 清洗 除锈 Of the medical instrument Cleaning In addition to rusty
  • 相关文献

参考文献3

二级参考文献15

  • 1贾淑芳,郭顺明.医院内感染的病原菌及耐药谱调查分析[J].上海医学检验杂志,1994,9(2):89-90. 被引量:8
  • 2陈昭斌.抗消毒剂细菌抗性机理的研究进展[J].现代预防医学,1994,21(4):241-243. 被引量:4
  • 3卫生部医政司医院感染监控协调小组.全国医院感染监控系统监测报告[J].中华流行病学杂志,1991,12(5):21-21.
  • 4陆德源.医学微生物学(第3版)[M].北京:人民卫生出版社,1990.27.
  • 5苑灿明 昊清华 等.苍山县村卫生室使用中消毒液微生物污染检测[J].中国消毒学杂志,1996,13(1):55-55.
  • 6薛广波.实用消毒学(第2版)[M].北京:人民军医出版社,1986.35-44.
  • 7卫生部医政司医院内感染监控协调小组,中华流行病学杂志,1991年,12卷,特5期,21页
  • 8苑灿明,中国消毒学杂志,1996年,13卷,1期,55页
  • 9陆德源,医学微生物学(第3版),1990年,95页
  • 10薛广波,实用消毒学(第2版),1986年,35页

共引文献616

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部